Non-invasive blood pressure (NiBP)

Start monitoring

The main steps can be undertaken at the parameter box.
1 Plug the hose connectors into the parameter box.
The rubber rings on the connectors can be moistened
slightly so that they can be inserted more easily.
Apply the appropriate cuff to the patient and connect
it to the hose. Ensure that all connections are tight.
Even slight leaks make measurement impossible.
Manual measurement
2 Press
.
The cuff is inflated and the cuff pressure indicated
constantly in the bar graph within the module.
3 The systolic, diastolic and
4 mean pressure are displayed when the measurement
is complete.
Incorrect measurements are repeated 5 seconds after
releasing the air. A third and last attempt is made after
30 seconds if the second measurement also proves
incorrect.
Automatic measurement
5 Time bar indicating the relative time elapsed between
two automatic measurements. The actual cuff
pressure is indicated here during the measurement.
¡
6 Time interval for automatic measurement.
7 Press
.
The LED in the key lights up and automatic measure-
ment starts. NiBP measurements are performed by
the parameter box at the set time intervals.
To interrupt measurement
2 Press
.
to interrupt measurement. The air is immediately
expelled from the cuff.