Testing The Pneumatic Circuit For Leaks; Testing Display Errors - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

Checking the NiBP pneumatic circuit for
leaks
Pull the supply hose of the sphygmomanometer cuff from
the twin hose and connect it to a pump-ball via a T piece.
The cuff (normal size) remains in the measuring circuit,
tightly rolled up. Then proceed as follows:
Open the venting valve in the measuring circuit.
Wait at least 30 seconds.
Press the »config« softkey.
Move the cursor frame horizontally across the screen
to »calibration« by turning the rotary control. Press
the rotary control to confirm the selection.
Move the cursor frame vertically to »more« by turning
the rotary control. Press the rotary control to confirm
the action.
Move the cursor frame vertically to »NiBP test« by
turning the rotary control. Press the rotary control to
confirm the selection.
The cuff pressure displayed on screen should be 0.
Wait 30 to 35 seconds, and then press the »AUTO«
key on the parameter box.
Increase the pressure to the test pressure value with
the pump-ball and wait at least 30 seconds for the
system to settle.
Measure the pressure drop.
If the overpressure relief valve responds, the para-
meter box will be vented automatically at 300 mmHg
± 5%. It will also be vented if the pressure falls below
15 mmHg. Repeat the test.
Quit the calibration mode by pressing the rotary
control after selecting »Check NiBP«. Deactivate the
»AUTO« key.
Measurement system testing
Detach the cuff supply hose from the twin hose and
connect to normal pressure instead.
Start the test mode as described above for
»Checking the pneumatic circuit for leaks«.
With the pump and venting valve at normal pressure,
the test pressure can now be set in the system.
Quit the test mode as described above for »Checking
the pneumatic circuit for leaks«.
Warning:
IPPV alarm limits
standard screen 1- adults
06-12-98
11:59
00:00
IPPV alarm limits
Anesth.Agent
Alarms
selection
default
man.
auto
CO
on
off
2
gas
SpO
Isoflurane
on
off
2
parameter-
The anesthetic
box
agent is automati-
on
off
cally selected by
HLM
on
off
agent detection.
Pmax
on
off
AGas Mix
on
off
06-12-98
11:59
Check system configuration. Please confirm !
00:00
Check operational readiness
Advisory:
Caution:
auto set
pat. alarm.
auto set
vent. al.
alarm
screen
config.
meter
config.
Calibration
Settings
?
iBP 1
mode
adult
neon.
?
iBP 2
alarm sound
1 2 3 4 5 6 7 8 9
ECG pulse tone
0 1 2 3 4 5 6 7 8 9
ECG pulse tone
more
SpO
modulated
on
off
2
curve speed
12,5 25
50
mm/s
colours
call standard
limits
info
para-
timer
start
135

Advertisement

Table of Contents
loading

Table of Contents