Dräger Cicero EM Instructions For Use Manual page 130

Integrated anaesthetic workstation
Table of Contents

Advertisement

Reassembling machine
Reassembling the parts
Fit flow sensor:
1 Unscrew expiration nozzle.
2 Slide flow sensor into housing.
1 Screw expiration nozzle back into place.
The sensor is automatically calibrated.
Fit microbial filter:
Fit a microbial filter with 22 mm fittings on both the
inspiration and expiration ports.
Fit temperature sensor (breathing gas):
(Optional) with Y-piece and hose clips.
The T-piece must be used for anaesthetic gas measure-
ment in this case.
3 Insert temperature sensor as far as possible into the
hole in the Y-piece.
After securing the Y-piece, align the sensor to face
upwards so that it remains free from condensation.
4 Route the sensor lead back to the machine with
hose clips along the inspiration hose of the
anaesthetic system.
Install bronchial aspirator
Position the bronchial aspirator on the mount on the
frame with two secretion collecting vessels.
5 Install hose between negative pressure outlet and
secretion collecting vessels.
6 Fit the handpiece (»fingertip«) on the aspiration hose.
7 Plug the aspiration hose with handpiece onto the inlet.
Secure the aspiration hose in the clamp.
130
2
1
3
D
6
5
7
4
4