Dräger Cicero EM Instructions For Use Manual page 51

Integrated anaesthetic workstation
Table of Contents

Advertisement

NiBP settings
(see also page 91)
time interval
Sets the time between measure-
ments.
active alarms
Determines which parameter is to
be monitored (systolic or diastolic
pressure).
initial pressure
Sets the starting pressure.
punction pressure
Sets the punction pressure.
The initial and punction pressure can be set separately
for adults and neonates. To do so, switch from »adult« to
»neon« and vice-versa under »Mode«.
unit
Selects the measuring units
interlock
When set to »on«, peripheral pulse
alarms are suppressed while the
sphygmomanometer cuff is being
inflated.
iBP locations
(See also page 94) With this function, measuring sites
can be preselected independently of the channel. The
following locations are available:
ART
Artery
AORTA
Aorta
A.Pulm.
Pulmonary artery
CVP
Central venous pressure
ICP
Intracranial pressure
?
Other location
The following parameters can be selected:
monitoring
Systolic, diastolic or mean pressure.
graph
abs
= fixed curve scale
morph = variable scale. For maximum
curve height.
amplitude
Scale of the curve.
Can only be used with fixed scales
(»abs«).
pulse
Switches pulse indication on/off.
iBP channels
defines input channels P1 and P2 of the parameter box:
measuring function
Switches the channel on/off.
location
Determines the location for this channel.
sensitivity
Selects the sensor sensitivity.
42.5 or 50 µV/V/mmHg
Configuring the system screen in »Standby«
Standby / Configuration
NiBP
time
interval
3
active alarms
sys
dia
initial pres-
sure adult
180 mmHg
punct. press.
adults
40
mmHg
units
mmHg kPa
interlock:
NiBP-SpO
on
off
2
NiBP-iBP
on
off
06-12-98
11:59
Select a menu position. Please confirm !
00:00
Change the settings. Please confirm again !
Standby / Configuration
ART
active alarms
sys
dia
graph
absol.
morph.
amplitude
160
pulse
on
off
AORTA
active alarms
sys
dia
graph
absol.
morph.
amplitude
160
pulse
on
off
mean
A. Pulm
active alarms
sys
dia
graph
absol.
morph.
amplitude
40
CVP
active alarms
mean
graph
absol.
morph.
amplitude
20
ICP
active alarms
mean
graph
absol.
morph.
amplitude
20
?
active alarms
sys
dia
mean
graph
absol.
morph.
amplitude
160
06-12-98
11:59
Select a menu position. Please confirm !
00:00
Change the settings. Please confirm again !
Standby / Configuration
iBP 1
function
OFF
ON
location
ART
sensitivity
42.5
50
iBP 2
function
OFF
ON
location
CVP
sensitivity
42.5
50
Select a menu position. Please confirm !
06-12-98
11:59
00:00
Change the settings. Please confirm again !
System screen functions
alarms inactive !
Settings
ECG
mode
adult
neon.
NiBP
alarm limits
iBP locations
parameter
iBP channels
screen
SpO
/ Pleth
acoustic
2
CO
/O
2
2
list entry
pEEG
transport
RS 232 (Medibus)
record (printer)
basic config.
alarms inactive !
Settings
ECG
mode
adult
neon.
NiBP
alarm limits
iBP locations
parameter
iBP channels
screen
SpO
/ Pleth
acoustic
2
CO
/O
2
2
list entry
pEEG
transport
RS 232 (Medibus)
record (printer)
basic config.
alarms inactive !
Settings
mode adult neon.
ECG
alarm limits
NiBP
iBP locations
parameter
iBP channels
screen
SpO
/ Pleth
2
acoustic
CO
/ O
2
2
list entry
pEEG
transport
RS 232 (Medibus)
record (printer)
basic config.
51