Ola Specific Commands - Agilent Technologies HP E5574A User Manual

Optical loss analyzer
Table of Contents

Advertisement

8.5
command
syntax
description
errors
example
command
syntax
description
extension
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
HP-IB Programming
OLA Specific Commands
OLA Specific Commands
This section describes the commands used to operate the OLA
remotely via its HP-IB interface.
The commands and queries appear in alphabetical order.
:ABORt
:ABORt
The abort command stops the automatic sweep of the polarization
controller. It is only applicable to the applications
• Powermeter,
• Powermeter Min/Max, and
• Return Loss.
The message "Wrong application for this command" is returned, if
the present application is not Powermeter, Powermeter Min/Max, or
Return Loss.
OUTPUT 724;":ABOR"
:INITiate
:INITiate[:IMMediate]
The initiate command starts the automatic sweep of the polarization
controller. It is only applicable to the applications
• Powermeter,
• Powermeter Min/Max, and
• Return Loss.
:IMM The ":IMMediate" extension may be present, but is
ignored.