Agilent Technologies HP E5574A User Manual page 117

Optical loss analyzer
Table of Contents

Advertisement

command
syntax
description
parameter
related commands
example
command
syntax
description
parameter
related commands
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
HP-IB Programming
IEEE Common Commands
*SAV
*SAV<wsp><location
With the SAVe command *SAV the instrument setting is stored in
RAM. You can store settings at locations 1 - 9. The scope of the
saved setting is identical to the standard setting (see *RST).
The RAM location: 1 ≤ location ≤ 9.
*RCL
OUTPUT 724;"*SAV 3"
*SRE
*SRE<wsp><value>
The Service Request Enable command *SRE sets bits in the
Service Request Enable register.
Any logical '1' in the enable register enables the corresponding bit
in the Status register.
The register is cleared at power-on. The *RST and *CLS
commands do not change the register.
The bit value for the register: 0 ≤ value ≤ 255.
Bit
Mnemonic
7 (MSB)
Operation Status
6
Request Status
5
Event Status Bit
4
Message Available
3
Questionable Status
2
Not used
1
Not used
0 (LSB)
Not used
Note: Bit 6 cannot be masked.
*SRE?
115
Decimal Value
128
64
32
16
8
0
0
0