Technical Data, Product Specifications
and Characteristics
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Appendix C. Specifications
Technical Data, Product Specifications and Characteristics
The HP E5574A Optical Loss Analyzer is produced to ISO 9001
international quality system standard as part of HP's commitment
to continually increasing customer satisfaction through improved
quality control.
Specifications describe the instrument's warranted performance.
Supplementary Performance Characteristics describe the
instrument's non-warranted typical performance.
The display may vary by ±1 count.
Notes for following table "Absolute Power Measurement
Specifications":
[1] At the following reference conditions:
• Power level 10 µW (-20 dBm), continuous wave (CW).
• Parallel beam, 3mm spot diameter on detector.
• Ambient temperature 23°C ±5K.
• At day of calibration (add 0.3% for aging over one year, add 0.6%
over two years).
• Spectral width of source <10nm.
[2] At the following operating conditions:
• Parallel beam, 3 mm spot diameter on detector or connectorized fiber
with NA ≤0.2.
• Ambient temperature within operating temperature range, non-
condensing.
• Within 1 year after calibration, add 0.3% for second year.
[3] Add ±0.007dB/dB between 10 and 27dBm.
[4] For single-mode fiber with NA ≤ 0.1, straight fiber end, 1250 –
1570 nm.
[5] with HP E5574A, setting optimized to DYNAMIC.
[6] with HP E5574A, setting optimized to PDL and with HP 8153A.