Agilent Technologies HP E5574A User Manual page 135

Optical loss analyzer
Table of Contents

Advertisement

response
errors
example
command
syntax
description
parameter
response
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
HP-IB Programming
OLA Specific Commands
The measured value in dB.
Note: The delay between sending the command and receiving a
response depends on the Window Size, the Window Mode and the
Average Time. Worst case:
Window Mode
Window Size
Average Time T avg
In this case the measurement will take 1000 s or approximately 17
minutes.
The message "Wrong application for this command" is returned, if
the present application is not MINMax.
The message "No head connected" is returned, if there is no optical
head attached to the specified channel.
The message "No valid result possible" is returned, if the value
cannot be measured.
OUTPUT 724;"SENS2:DATA? MINM"
ENTER 724;A$
:SENSe:DATA? PDL
:SENSe[1|2]:DATA?<wsp>PDL
This query returns the PDL value. It is only applicable to the
applications
• Polarization Dependent Loss,
• PDL/Insertion Loss, and
• Polarization Dependent Coupler Test.
1|2
1
Reads head A
2
Reads head B
Default is 1.
The measured value in dB.
133
"refresh"
1000
1 s