HP-IB Programming
IEEE Common Commands
*OPC, *WAI
OUTPUT 724;"*CLS;*ESE 1;*SRE 32"
OUTPUT 724;"*OPC?"

ENTER 724;A$

*OPT?
*OPT?
The OPTions query *OPT? returns a string with the options
installed in the OLA. There are two fields, separated by commas.
The two fields are: <Laser>,<Connector>
Laser:
1310nm, 1550nm, or 1310nm/1550nm
Connector:
Bare Fiber, Straight Contact, or Angled Contact
*IDN?
DIM A$[60]
OUTPUT 724;"*OPT?"
ENTER 724;A$
*RCL
*RCL<wsp><location>
The ReCaLl instrument setting command *RCL changes the
instrument setting to that stored at the internal RAM location
specified.
Location 0 contains the default setting for the instrument. For
information on the default setting, see *RST. For information on
how to store individual settings at locations 1 to 9, see *SAV.
The RAM location: 0 ≤ location ≤ 9.
*SAV
OUTPUT 724;"*RCL"
113