HP-IB Programming
IEEE Common Commands
*RST
*RST
The ReSeT command *RST sets the instrument to the default
setting stored in ROM. Pending *OPC/*OPC? actions are
cancelled.
The ReSeT command acts as follows:
• selects the default laser (1310 nm if installed, 1550 nm else)
• switches the laser off
• activates the "Select Application" menu
• sets the average time to 200 ms
• sets the reference power to -10 dBm
• sets RL ref to 14.7 dB
• sets the unit for power to Watt
• sets the unit for SR in CT and PDCT measurement to %
• stops the automatic sweep of the polarization controller
• sets the scan rate of the polarization controller to 5
• activates head A
The instrument is placed in the idle state waiting for a command.
The *RST command clears the input queue.
The following are not changed:
• HP-IB (interface) state
• Instrument interface address
• Output queue
• Service Request Enable register (SRE)
• Standard Event Status Enable register (ESE)
OUTPUT 724;"*RST"