HP-IB Programming
OLA Specific Commands
Note:
• The Stability application cannot be started from the "Select
Application" menu application, from the Powermeter
application, or from the Powermeter Min/Max application.
• To return to the Powermeter application from the Min/Max
application, you have to restart the Powermeter application by
first sending the command :SENS:FUNC MAIN and then
sending the command :SENS:FUNC POW.
• The command :SENS:FUNC MAIN stops any active application
and returns to the "Select Application" menu.
The message "Wrong application for this command" is returned, if
you specify STAB and the present application is Powermeter,
Powermeter Min/Max or none.
The same message is returned, if you specify MINM and the
present application is not Powermeter.
:SENSe:FUNCtion?, :SENSe:FUNCtion:STATe?
OUTPUT 724;":SENS:FUNC MAIN"
OUTPUT 724;":SENS:FUNC PDL"
:SENSe:FUNCtion?
:SENSe[1|2]:FUNCtion[:ON]?
This query returns the application currently active as a string. It can
be used any time.
1|2 The channel parameter 1 or 2 may be present, but is ignored.
:ON The ":ON" extension may be present, but is ignored.