Appendix D. Performance Test
Test IV. Linearity and Accuracy
5 Alter the attenuation until the HP 8153A displays –17.4dBm.
a At the Attenuator:
Press A
and use the modify cursor keys to alter the
TT
attenuation.
6 On the HP 8153A:
a Press D
–>R
ISP
EF
b Press U
to select the –10 dBm range.
P
7 Note the deviation, displayed in dB, as R1 in the test record.
8 In the –10 dBm range.
a Set the display to show results in dBm.
b Alter the attenuation until the DUT displays –7.4 dBm.
c
Press D
–>R
ISP
EF
d Select the 0 dBm range.
e
Note the deviation, displayed in dB, as R2 on the test record.
9 Repeat the steps a. to e. at all ranges and level settings shown in
the test record.
10 Calculate the non-linearity using the formula in the test record.