related commands
example
command
syntax
description
parameter
response
errors
example
command
syntax
description
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
HP-IB Programming
OLA Specific Commands
:SENSe:POWer:MEASuring:MODe, :SENSe:POWer:UNIT
OUTPUT 724;":SENS1:DATA? POW"
ENTER 724;A$
:SENSe:DATA? RL
:SENSe[1|2]:DATA?<wsp>RL
This query returns the Return Loss value. It is only applicable to the
Return Loss application.
1|2
1
Reads head A
2
Reads head B
Default is 1.
The measured value in dB.
The message "Wrong application for this command" is returned, if
the present application is not Return Loss.
The message "No head connected" is returned, if there is no optical
head attached to the specified channel.
The message "No valid result possible" is returned, if the value
could not be measured.
OUTPUT 724;":SENS1:DATA? RL"
ENTER 724;A$
:SENSe:DATA? SR
:SENSe[1|2]:DATA? SR
This query returns either the Splitting Ratio SR or the Polarization
Dependent Splitting Ratio PDSR, depending on the application
currently active. It is only applicable to the applications
• Coupler Test, and
• Polarization Dependent Coupler Test.
135