Agilent Technologies HP E5574A User Manual page 128

Optical loss analyzer
Table of Contents

Advertisement

errors
example
command
syntax
description
parameters
errors
example
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
HP-IB Programming
OLA Specific Commands
The message "Wrong application for this command" is returned, if
the present application is not Powermeter, Powermeter Min/Max, or
Return Loss.
OUTPUT 724;":INIT"
:PADDle:POSition
:PADDle[1|2|3|4]:POSition<wsp><value|MIN|MAX>
The paddle positioning command sets the specified paddle (coil) of
the polarization controller to the desired angle position. It is only
applicable to the applications
• Powermeter,
• Powermeter Min/Max, and
• Return Loss.
Note: This command is only executed when the automatic sweep is
off (see ":ABORt").
1|2|3|4 The number of the paddle which is to be set.
Default is 1.
value The value for the angle: 0 ≤ value ≤ 999).
Value
Angle
0
249
45°
499
90°
749
135°
999
179°
MIN|MAX Instead of a value you can enter the keyword MIN (=
0°) or MAX (= 180°) respectively.
The message "Wrong application for this command" is returned, if
the present application is not Powermeter, Powermeter Min/Max, or
Return Loss.
OUTPUT 724;"ABORT;:PADD2:POS 355"
126

Advertisement

Table of Contents
loading