HP-IB Programming
OLA Specific Commands
:SENSe:POWer:OPTimize?
:SENSe[1|2]:POWer:OPTimize?
1|2 The channel parameter 1 or 2 may be present, but is ignored.
This query returns the current optimization status of the OLA.
Note: The optimization affects only the temperature controlled
heads of type HP 81521B opt. 001.
A boolean value:
0
The OLA is optimized for PDL measurement
1
The OLA is optimized for maximal sensitivity
:SENSe:POWer:OPTimize
OUTPUT 724;":SENS:POW:OPT?"
ENTER 724;A$
:SENSe:POWer:REFerence
:SENSe[1|2]:POWer:REFerence<wsp><value>
This command sets the reference reflection RL ref to the specified
value. It is only applicable to the Return Loss application.
Note: This is the only reference, which is not measured.
1|2 The channel parameter 1 or 2 may be present, but is ignored.
Value The open end reflection in dB (usually 14.7).
:SENSe:POWer:REFerence?,
SENSe:POWer:REFerence:DISPlay:PPAR,
:SENSe:POWer:REFerence:DISPlay:PREF
The message "Wrong application for this command" is returned, if
the present application is not Return Loss.
OUTPUT 724;":SENS:POW:REF 14.7"