HP-IB Programming
OLA Specific Commands
The message "Wrong application for this command" is returned, if
you send DBM|W|0|1 and the current application is not
Powermeter.
The same message is returned, if you send PERCent|DB|2|3 and the
current application is not Coupler Test or Polarization Dependent
Coupler Test.
:SENSe:POWer:UNIT?, :SENSe:DATA? POW, :SENSe:DATA?
CR, :SENSe:DATA? SR
OUTPUT 724;":SENS:FUNC POW"
OUTPUT 724;":SENS:POW:UNIT W"
:SENSe:POWer:UNIT?
:SENSe[1|2]:POWer:UNIT?
This query returns the current setting of the measurement unit. It is
only applicable to the applications
• Coupler Test,
• Polarization Dependent Coupler Test, and
• Powermeter.
1|2 When the Powermeter application is active:
1
Reads head A
2
Reads head B
Default is 1.
When the Coupler Test or PD Coupler Test application is active:
The channel parameter 1 or 2 may be present, but is ignored.
An integer number. When the Powermeter application is active:
0
Measurement unit dBm
1
Measurement unit Watt (or mW)
3
Measurement unit dB (channel mode is set for relative
measurements)