HP-IB Programming
OLA Specific Commands
A boolean value:
0
Application is inactive
1
Application is active
:SENSe:FUNCtion, :SENSe:FUNCtion?
OUTPUT 724;":SENS:FUNC:STAT? 3"
ENTER 724;A$
:SENSe:POWer:ATIMe
:SENSe[1|2]:POWer:ATIMe<wsp><value>[unit]
This command sets the time interval used for averaging power
measurements. It is only applicable to the applications
• Powermeter, and
• Insertion Loss.
1|2 The channel parameter 1 or 2 may be present, but is ignored.
The setting affects both channels.
Unit s = seconds, ms = milliseconds. Default is s.
Value If unit is "s" or default: Choice of 0.02, 0.2, 1.
If unit is "ms": Choice of 20, 200, 1000.
Default is 0.2 s. If you transmit a different value, it is rounded to
one of the three values listed above.
The message "Wrong application for this command" is returned, if
the present application is not Powermeter or Insertion Loss.
:SENSe:POWer:ATIMe?
OUTPUT 724;":SENS:POW:ATIM 20ms"