command
syntax
description
parameter
response
related commands
example
command
syntax
description
parameter
response
example
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
HP-IB Programming
OLA Specific Commands
:SENSe:POWer:CALC:MODe?
:SENSe[1|2]:POWer:CALC:MODe?
This query returns the current setting of the measurement and
display mode for the polarization dependent applications. It can be
used any time.
1|2 The channel parameter 1 or 2 may be present, but is ignored.
A boolean value:
0
Refresh
1
Average
:SENSe:POWer:CALC:MODE
OUTPUT 724;":SENS:POW:CALC:MOD?"
ENTER 724;A$
:SENSe:POWer:HEAD?
:SENSe[1|2]:POWer:HEAD?
This query returns information about the optical head connected to
the specified channel. It can be used any time.
1|2
1 Reads head A
2 Reads head B
Default is 1.
An integer number:
0
There is no head attached to the specified channel
1
There is a head attached to the specified channel
2
A low PDL head is attached to the specified channel
3
A low PDL head with temperature control is attached to the
specified channel
OUTPUT 724;":SENS2:POW:HEAD?
ENTER 724;A$