Appendix D. Performance Test
Test V. PDL/PDG uncertainty (#020 only)
Test V. PDL/PDG uncertainty (#020 only)
The absolute PDL/PDG uncertainty is the maximum value of
several PDL measurement when changing the alignment of the
PDL generating components.
1 Set up and switch on the equipment:
a Connect the equipment as shown in Figure D-6.
(Set-up for performing the test with Head Input A. For
performing the test with Head Input B connect the (second)
HP 81521B#001 Head to Head Input B.)
HP E5574 Optical Loss Analyzer
Head A
Head B
Input
HP 81521B #001
Optical Head
Optical Adapter
HP 81000FA
connecting fiber

PDL/PDG Uncertainty Test Setup

b Ensure that the instruments have warmed up and that the
laser has settled.
2 Set up the OLA:
a Optimize the OLA for PDL measurement:
Press P
RESET
b Select Polarization Dependent Loss Application:
Press A
PPL
3 Zero the OLA:
206
Output
and then press O
PTIMIZE
, select Pol Depend. Loss and then press
to change to PDL.