HP-IB Programming
OLA Specific Commands
:SENSe:POWer:WAVelength?
:SENSe[1|2]:POWer:WAVelength?
This query returns the current wavelength setting.
1|2 The channel parameter 1 or 2 may be present, but is ignored.
The wavelength is valid for both channels.
The wavelength is returned as a floating point number in
exponential format, e.g. 1.3E-6. The unit is meter. The unit is not
included in the response.
:SENSe:POWer:WAVelength
OUTPUT 724;":SENS:POW:WAV?"
ENTER 724;A$
:SENSe:POWer:WINDow:MODe
:SENSe[1|2]:POWer:WINDow:MODe<wsp><mode>
This command sets the mode of the measurement window used for
calculating the Min/Max value. It is only applicable to the
Powermeter Min/Max application.
1|2 The channel parameter 1 or 2 may be present, but is ignored.
The setting affects both channels.
Mode There are three modes available. You enter one of the
following keywords:
SLIDing
The window is filled once. After that, the Min/
Max value is updated from sample to sample.
REFResh
The Min/Max value is not updated until the
window has been refilled completely.
CONTinuous
The Min/Max value is calculated as the
difference between the present and the previous
value.
The message "Wrong application for this command" is returned, if
the present application is not Powermeter Min/Max.
:SENSe:POWer:WINDow:MODe?
154