CAUTI ON
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Appendix E. Cleaning Procedure
Cleaning Procedures
Cleaning Connector Interfaces
Be careful when using pipe-cleaners, as the core and the bristles of the pipe-
cleaner are hard and can damage the interface.
Do not use pipe-cleaners on optical head adapters, as the hard core of
normal pipe cleaners can damage the bottom of an adapter.
Preferred Procedure
1 Clean the interface by pushing and pulling a new, dry pipe-
cleaner into the opening. Rotate the pipe-cleaner slowly as you
do this.
2 Then clean the interface by rubbing a new, dry cotton-swab over
the surface using a small circular movement.
3 Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure particularly when there is greasy dirt on the
interface:
1 Moisten a new pipe-cleaner with isopropyl alcohol.
2 Clean the interface by pushing and pulling the pipe-cleaner into
the opening. Rotate the pipe-cleaner slowly as you do this.
3 Moisten a new cotton-swab with isopropyl alcohol.
4 Clean the interface by rubbing the cotton-swab over the surface
using a small circular movement.
5 Using a new, dry pipe-cleaner, and a new, dry cotton-swab
remove the alcohol, any dissolved sediment and dust.
6 Blow away any remaining lint with compressed air.