related commands
errors
example
command
syntax
description
parameter
response
related commands
example
command
syntax
description
parameter
related commands
example
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
HP-IB Programming
OLA Specific Commands
:SENSe:POWer:WINDow:SIZe?
The message "Wrong application for this command" is returned, if
the present application is not Powermeter Min/Max.
OUTPUT 724;":SENS:POW:WIND:SIZ 500"
:SENSe:POWer:WINDow:SIZe?
:SENSe[1|2]:POWer:WINDow:SIZe?
This query returns the dimension of the window used for collecting
the samples for the Min/Max determination. It is only applicable to
the Powermeter Min/Max application.
1|2 The channel parameter 1 or 2 may be present, but is ignored.
The setting affects both channels.
The number of samples which constitutes the window: 50, 100,
500, or 1000.
:SENSe:POWer:WINDow:SIZe
OUTPUT 724;":SENS:POW:WIND:SIZ?"
ENTER 724;A$
:SOURce:POWer:STATe
:SOURce:POWer:STATe<wsp><boolean>
This command switches the built-in laser source on or off. It is
applicable any time.
Boolean A boolean variable: OFF|ON|0|1. You can enter either
the function literally or the corresponding number:
OFF
0 Switches the internal laser source off.
ON
1 Switches the internal laser source on.
Note: Any non-zero number switches the internal laser source on.
:SOURce:POWer:STATe?
OUTPUT 724;":SOUR:POW:STAT ON"