Taking Standard Loss Measurements
Measuring the Return Loss
8 Press
C
ALIBRATE

Return Loss Calibration Display

9 Press
D
–>P
ISP
optical output and received by the optical head is stored as the
reference power.
10 Wind the optical output fiber cable close to the open end in seven
tight loops around a pen.
This almost eliminates the reflection from the open fiber end.
The remaining optical power is caused by so-called parasitic
reflections within the test setup. It should be as small as
possible (below –50 dBm).
11 Press
D
–>P
ISP
power.
12 Unwind the optical output fiber cable.
13 Press
P
REVIOUS
The display shows the open-end reflection.
You now are ready to attach your DUT.