Agilent Technologies HP E5574A User Manual page 159

Optical loss analyzer
Table of Contents

Advertisement

command
syntax
description
response
related commands
example
command
syntax
description
parameter
related commands
example
command
syntax
description
response
related commands
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
HP-IB Programming
OLA Specific Commands
:SOURce:POWer:STATe?
:SOURce:POWer:STATe?
This query returns the current state of the internal laser source.
A boolean value:
0
The internal source is off.
1
The internal source is on.
:SOURce:POWer:STATe
OUTPUT 724;":SOUR:POW:STAT?"
ENTER 724;A$
:SOURce:POWer:WAVelength
:SOURce:POWer:WAVelength<wsp><source>
This command switches the wavelength of the built-in laser source.
It is only applicable to an instrument equipped with the dual source
option. It can be used any time. It is ignored, if the OLA has only a
single source.
Source You enter one of the following keywords:
LOWer
Switches to 1310 nm
UPPer
Switches to 1550 nm
:SOURce:POWer:WAVelength?
OUTPUT 724;"SOUR:POW:WAV UPP"
:SOURce:POWer:WAVelength?
:SOURce:POWer:WAVelength?
This query returns the current wavelength of the internal source.
The wavelength is returned as a floating point number in
exponential format, e.g. 1.55E-6. The unit is meter. The unit is not
included in the response.
:SOURce:POWer:WAVelength
157