HP-IB Programming
OLA Specific Commands
OUTPUT 724;":SENS:POW:WIND:MOD REFR"
:SENSe:POWer:WINDow:MODe?
:SENSe[1|2]:POWer:WINDow:MODe?
This query returns the current mode of the window used for
collecting the samples for the Min/Max determination. It is only
applicable to the Powermeter Min/Max application.
1|2 The channel parameter 1 or 2 may be present, but is ignored.
The setting affects both channels.
An integer value:
0
Sliding window
1
Refresh mode
2
Continuous mode
The message "Wrong application for this command" is returned, if
the present application is not MINMax.
:SENSe:POWer:WINDow:MODe
OUTPUT 724;":SENS:POW:WIND:MOD?"
ENTER 724;A$
:SENSe:POWer:WINDow:SIZe
:SENSe[1|2]:POWer:WINDow:SIZe<wsp><value>
This command specifies the dimension of the window used for
collecting the samples for the Min/Max determination. It is only
applicable to the Powermeter Min/Max application.
1|2 The channel parameter 1 or 2 may be present, but is ignored.
The setting affects both channels.
Value The number of samples used to determine the minimum
and maximum value: 50|100|500|1000.
Note: If you specify a number which is different from those listed
above, the window is set to the size which comes closest. For
instance, 74 sets 50, 75 sets 100 samples.
155