Agilent Technologies HP E5574A User Manual page 145

Optical loss analyzer
Table of Contents

Advertisement

command
syntax
description
parameters
errors
related commands
examples
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
HP-IB Programming
OLA Specific Commands
:SENSe:POWer:MEASuring:MODe
:SENSe[1|2]:POWer:MEASuring:MODe<wsp><mode>
This command sets the power measurement mode for the specified
channel. It is only applicable to the Powermeter application.
Note: The command which returns a result is :SENS:DATA? POW.
1|2
1 Sets head A
2 Sets head B
Default is 1.
Mode ABS|REL1|REL2|0|1|2. You can enter either the mnemonic
or the corresponding number.
If you choose head A:
ABS
0
Absolute power measurement
REL1
1
Relative measurement A/Ref
REL2
2
Relative measurement A/B
If you choose head B:
ABS
0
Absolute power measurement
REL1
1
Relative measurement B/Ref
REL2
2
Relative measurement B/A
The message "Wrong application for this command" is returned, if
the current application is not Powermeter.
:SENSe:POWer:MEASuring:MODe?, :SENS:DATA? POW
The following examples yield the same instrument setting:
OUTPUT 724;":SENS1:POW:MEAS:MOD ABS"
OUTPUT 724;":SENS1:POW:MEAS:MOD 0"
143