Page 2
DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data. Certification Agilent Technologies, Inc. certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies, Inc. further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members.
Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies, Inc. products. For information about these agreements and for other assistance, contact Agilent. Refer to “Contacting Agilent” on page 2-13. Safety and Regulatory Information The safety and regulatory information pertaining to this product is located in Chapter 1, “Safety and Regulatory Information.”...
Page 4
Documentation Map The online Help files are embedded in the analyzer, offering quick reference to programming and user documentation. From the Help drop-down menu, you can access the Help system in five different languages. Also, you can view the Analyzer Product Overview multimedia presentation and access the analyzer’s Web page.
Page 10
Contents Removing and Replacing the 30-dB Fixed Attenuators (E8362B only) ....7-44 Removing and Replacing the A23 and A24 Detectors (E8363B and E8364B only) ..7-46 Removing and Replacing the A25 and A26 Test Port Couplers .
Safety and Regulatory Information PNA Series Microwave Network Analyzers Information in This Chapter E8362B, E8363B, E8364B Information in This Chapter This chapter provides safety information that will help protect you and your network analyzer. It also contains information that is required by various government regulatory agencies.
PNA Series Microwave Network Analyzers Safety and Regulatory Information E8362B, E8363B, E8364B Safety Symbols Safety Symbols The following safety symbols are used throughout this manual. Familiarize yourself with each of the symbols and its meaning before operating this instrument. CAUTION Caution denotes a hazard.
Safety and Regulatory Information PNA Series Microwave Network Analyzers General Safety Considerations E8362B, E8363B, E8364B CAUTION This instrument has autoranging line voltage input. Be sure the supply voltage is within the specified range. CAUTION This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 61010-1:2001 and 664 respectively.
Page 15
PNA Series Microwave Network Analyzers Safety and Regulatory Information E8362B, E8363B, E8364B General Safety Considerations WARNING For continued protection against fire hazard, replace line fuse only with same type and rating. The use of other fuses or material is prohibited.
Safety and Regulatory Information PNA Series Microwave Network Analyzers Electrostatic Discharge Protection E8362B, E8363B, E8364B Electrostatic Discharge Protection Protection against electrostatic discharge (ESD) is essential while removing assemblies from or connecting cables to the network analyzer. Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged.
PNA Series Microwave Network Analyzers Safety and Regulatory Information E8362B, E8363B, E8364B Regulatory Information Regulatory Information This section contains information that is required by various government regulatory agencies. Instrument Markings The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
If the battery on the A15 CPU board assembly needs to be disposed of, dispose of it in accordance with your country’s requirements. If required, you may return the battery to Agilent Technologies for disposal. Refer to “Contacting Agilent” on page 2-13 assistance.
General Product Information PNA Series Microwave Network Analyzers Information in This Chapter E8362B, E8363B, E8364B Information in This Chapter Chapter Two at-a-Glance Section Title Summary of Content Start Page Cleaning instructions for the external surfaces of your analyzer. Maintenance Page 2-3 Information about electrical maintenance of your analyzer.
Upgrades Available” on page 2-7. Individual upgrade kits are also available and can be ordered by part number. Refer to “E8362B Upgrade Kits” on page 6-81 “E8363B, E8364B Upgrade Kits” on page 6-81 for a complete list of available kits.
This upgrade—E8362BU-040—provides the additional hardware and software necessary to make frequency measurements up to 40 GHz with your E8362B PNA. If you have a standard E8362B PNA, then upgrade E8362BU-040 provides the additional hardware and software necessary to make frequency measurements up to 40 GHz.
PNA Series Microwave Network Analyzers General Product Information E8362B, E8363B, E8364B Analyzer Options Available Option 081, External Reference Switch This upgrade adds a solid-state internal RF transfer switch in the R1 reference receiver path. The switch allows the instrument to easily switch between standard S-parameter (non-frequency offset) measurements and frequency offset measurements such as relative phase or absolute group delay that require an external reference mixer.
General Product Information PNA Series Microwave Network Analyzers Analyzer Options Available E8362B, E8363B, E8364B Option 1CM, Rack Mount Flange Kit for Instruments without Handles Option 1CM provides a rack mount kit that contains a pair of flanges, rack mount rails, and the necessary hardware to mount the analyzer in an equipment rack with 482.6 mm...
Configurable test set (Option 014) E8362/3/4BU-014 Receiver attenuators (Option 016) E8362/3/4BU-016 Extended frequency range to 40 GHz for E8362B without Option 014 (Option 040) E8362BU-040 Extended frequency range to 40 GHz for E8362B with Option 014 (Option 041) E8362BU-041 Extended frequency range to 50 GHz for E8362B without Option 014 (Option 050)
General Product Information PNA Series Microwave Network Analyzers Analyzer Accessories Available E8362B, E8363B, E8364B Analyzer Accessories Available CD-RW Drive This accessory is an external read/write CD drive with a USB cable and can be ordered as model number N4688A. USB Hub This accessory is a 4-port USB hub for connecting additional USB peripheral devices and can be ordered as model number N4689A.
PNA Series Microwave Network Analyzers General Product Information E8362B, E8363B, E8364B Required Service Test Equipment Required Service Test Equipment Recommended Alternate Critical Specifications Model or Part Model or Part Equipment Number Number Test Instruments and Software Freq: 10 MHz to 20 GHz...
Page 28
General Product Information PNA Series Microwave Network Analyzers Required Service Test Equipment E8362B, E8363B, E8364B Recommended Alternate Critical Specifications Model or Part Model or Part Equipment Number Number Calibration and Verification Kits Type-N calibration kit Freq: 10 MHz to 20 GHz...
Page 29
PNA Series Microwave Network Analyzers General Product Information E8362B, E8363B, E8364B Required Service Test Equipment Recommended Alternate Critical Specifications Model or Part Model Equipment Number Number Tools Extender board E8356-60021 None T-8 TORX driver 0.6 N-m (5 in-lb) setting 0.5, 0.8, and 1.0 N-m...
“Contacting Agilent” • “Shipping Your Analyzer to Agilent for Service or Repair” Service and Support Options The analyzer’s standard warranty is a one-year return to Agilent Technologies service warranty. NOTE There are many other repair and calibration options available from the Agilent Technologies support organization.
Agilent for repair. If you wish to send your network analyzer to Agilent Technologies for service or repair: • Include a complete description of the service requested or of the failure and a description of any failed test and any error message.
Page 32
General Product Information PNA Series Microwave Network Analyzers Agilent Support, Services, and Assistance E8362B, E8363B, E8364B 2-14 Service Guide E8364-90026...
Tests and Adjustments PNA Series Microwave Network Analyzers Information in This Chapter E8362B, E8363B, E8364B Information in This Chapter This chapter contains procedures to help you check, verify, and adjust your PNA series network analyzer. • The checks verify the operation of the assemblies in your analyzer.
Page 35
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Information in This Chapter Section Title Summary of Content Start Page Performing the operator’s check. Checking your test cables. Preliminary Checks Page 3-11 Perform these checks before performing system verification.
Tests and Adjustments PNA Series Microwave Network Analyzers Before You Begin E8362B, E8363B, E8364B Before You Begin Before checking, verifying, or adjusting the analyzer, refer to the following paragraphs to: • make sure the operating environment is within its requirements •...
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Before You Begin Review the Principles of Connector Care Proper connector care and connection techniques are critical for accurate and repeatable measurements. Refer to Table 3-1 for tips on connector care.
Tests and Adjustments PNA Series Microwave Network Analyzers About System Verification and Performance Tests E8362B, E8363B, E8364B About System Verification and Performance Tests The performance of the network analyzer is specified in two ways: system specifications, and instrument specifications. It is the end user’s responsibility to determine which set of specifications is applicable to their use of the PNA.
3-10. Certificate of Calibration Agilent Technologies will issue a certificate of calibration upon successful completion of system verification or completion of the performance tests. The certificate of calibration will apply to the “system” (analyzer, calibration kit, test cables, and any necessary adapters) if the system verification procedure is used to confirm the system specifications.
Page 40
Tests and Adjustments PNA Series Microwave Network Analyzers About System Verification and Performance Tests E8362B, E8363B, E8364B inaccurate measurements if you then use the analyzer in an application requiring additional capabilities. Service Guide E8364-90026...
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B ANSI/NCSL Z540–1–1994 Verification ANSI/NCSL Z540–1–1994 Verification To meet the criteria for ANSI/NCSL Z540-1-1994, perform the preliminary checks and either system verification or performance tests without stopping to repair or adjust .
Tests and Adjustments PNA Series Microwave Network Analyzers Non-ANSI/NCSL Z540–1–1994 Verification E8362B, E8363B, E8364B Non-ANSI/NCSL Z540–1–1994 Verification For non-ANSI/NCSL Z540-1-1994, perform the preliminary checks and either the system verification or performance tests while stopping to troubleshoot. Refer to Figure 3-2 for test flow.
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Preliminary Checks Preliminary Checks Preliminary checks include the following: • “The Operator’s Check” on page 3-11 The operator’s check tests the network analyzer’s basic functionality of the source, switch, step attenuators, and receivers.
Page 44
Tests and Adjustments PNA Series Microwave Network Analyzers Preliminary Checks E8362B, E8363B, E8364B 3. Click Begin. 4. If shorts and opens are not connected to all ports, you will be prompted to connect them as they are needed. 5. The result of the operator’s check will be shown as a PASS or FAIL next to each test...
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Preliminary Checks The Test Port Cable Checks A faulty test port cable can cause a failure in the verification test. The following checks are not required, but are recommended to verify the performance of the test port cable.
Page 46
Tests and Adjustments PNA Series Microwave Network Analyzers Preliminary Checks E8362B, E8363B, E8364B Cable Return Loss Check 1. Press Preset. 2. Perform a one-port calibration on Port 1, 1-Port Reflection. Refer to the embedded help in the analyzer if necessary.
Page 47
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Preliminary Checks Cable Insertion Loss Check 1. With the test port cable still connected to Port 1, connect a short to the other end of the cable. 2. From the Marker menu, click Marker Search. In the Marker Search dialog box, in the Search Type box, select Minimum.
Page 48
Tests and Adjustments PNA Series Microwave Network Analyzers Preliminary Checks E8362B, E8363B, E8364B Cable Magnitude and Phase Stability Check 1. With the test port cable still connected to Port 1, connect a short to the other end of the cable.
Page 49
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Preliminary Checks Figure 3-6 Typical Cable Magnitude and Phase Stability Response If the Cable Magnitude and Phase Stability Check Fails 1. Clean the cable and devices and torque connections to specifications (refer to “Review...
Page 50
Tests and Adjustments PNA Series Microwave Network Analyzers Preliminary Checks E8362B, E8363B, E8364B Cable Connector Repeatability Check NOTE The connector repeatability measurement should be done at the test port as well as at the end of the test port cable.
Page 51
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Preliminary Checks 12. For the Port 2 Check, connect the cable (with the load attached) to Port 2 and repeat steps 2 through 11. Figure 3-7 Typical Cable Connector Repeatability Response If the Cable Connector Repeatability Check Fails 1.
Tests and Adjustments PNA Series Microwave Network Analyzers System Verification E8362B, E8363B, E8364B System Verification System verification is used to verify system-level, error-corrected uncertainty limits for network analyzer measurements. The verification procedure is automated and is contained in the firmware of the analyzer.
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B System Verification Measurement Uncertainty Measurement uncertainty is defined as the sum of: • the residual systematic (repeatable) errors, and • the random (non-repeatable) errors in the measurement system after calibration.
Tests and Adjustments PNA Series Microwave Network Analyzers System Verification E8362B, E8363B, E8364B Measurement Traceability To establish a measurement traceability path to a national standard for a network analyzer system, the overall system performance is verified through the measurement of devices that have a traceable path.
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B System Verification Performing System Verification The following verification procedure is automated by the analyzer firmware. For each verification device, the analyzer reads a file from the verification disk and sequentially measures the magnitude and phase for all four S-parameters.
Page 56
Tests and Adjustments PNA Series Microwave Network Analyzers System Verification E8362B, E8363B, E8364B Precautions for Handling Airlines The verification kit includes airlines. Follow the steps below to ensure good measurement techniques. • Be very careful not to drop the airline’s center or outer conductor. Irreparable damage will result if these devices are dropped.
Page 57
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B System Verification System Verification Procedure 1. Connect a cable to the analyzer as shown in Figure 3-11. If you desire printed test outputs, connect a printer to the analyzer. For the printer, ensure that the correct driver is loaded and the printer is defined as the default printer.
Page 58
PNA Series Microwave Network Analyzers System Verification E8362B, E8363B, E8364B 4. In the Calibration Kit box, select the calibration kit that is being used by clicking on it. The corresponding verification kit to use is selected for you and displayed in the Verification Kit box.
Page 59
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B System Verification If the System Fails the Verification Test Inspect all connections. Do not remove the cable from the analyzer test port. IMPORTANT This will invalidate the calibration that you performed earlier.
Page 60
Tests and Adjustments PNA Series Microwave Network Analyzers System Verification E8362B, E8363B, E8364B Interpreting the Verification Results Figure 3-15 shows an example of typical verification results with Tabular Data selected in the Printer Output area of the System Verification dialog box.
Page 61
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B System Verification Figure 3-16 shows an example of typical verification results with Measurement Plots selected in the Printer Output area of the System Verification windows. The printed graphical results show the following: •...
Tests and Adjustments PNA Series Microwave Network Analyzers Option H11, IF Access, Operation Verification E8362B, E8363B, E8364B Option H11, IF Access, Operation Verification To verify that the Option H11, IF Access, is functioning properly: 1. Download and install the verification program from the Web: a.
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Performance Tests (Agilent N7840A Software Package) Performance Tests (Agilent N7840A Software Package) The Agilent N7840A software package verifies the electrical performance of your E8364B PNA series microwave network analyzer. The software automatically configures your analyzer to execute the performance tests.
Tests and Adjustments PNA Series Microwave Network Analyzers Performance Tests (Agilent N7840A Software Package) E8362B, E8363B, E8364B Source Maximum Power Output Test Function of the Test: To confirm the maximum source output power of your network analyzer over its full frequency range.
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Performance Tests (Agilent N7840A Software Package) Frequency Accuracy Test Function of the Test: To verify the frequency accuracy and range of the analyzer’s source output. Specification Tested: Test Port Output–CW Accuracy Equipment Used: A frequency counter, a test cable, and adapters.
Tests and Adjustments PNA Series Microwave Network Analyzers Performance Tests (Agilent N7840A Software Package) E8362B, E8363B, E8364B Receiver Compression Test Function of the Test: To measure the compression at the analyzer’s specified maximum power level for the receivers. Specification Tested: Test Port Input–Maximum Test Port Input Level Equipment Used: The compression test set (Z5623A Option K01), a power meter, power sensors, two test cables, and a calibration kit.
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Performance Tests (Agilent N7840A Software Package) Noise Floor Test Function of the Test: To measure the absolute power level of the noise floor for the analyzer’s receivers. Specification Tested: Test Port Input–Test Port Noise Floor Equipment Used: A power meter, power sensor, a calibration kit, and a test cable.
Tests and Adjustments PNA Series Microwave Network Analyzers Performance Tests (Agilent N7840A Software Package) E8362B, E8363B, E8364B Calibration Coefficient Test Function of the Test: To verify the uncorrected calibration coefficients of your analyzer. The calibration coefficients are measured in forward and reverse direction. Refer to “Error...
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Performance Tests (Agilent N7840A Software Package) System Crosstalk Test Function of the Test: To measure the level of signal leakage between the analyzer’s test ports. Specification Tested: Test Port Input–Crosstalk Equipment Used: A calibration kit and a test cable.
Tests and Adjustments PNA Series Microwave Network Analyzers Performance Tests (Agilent N7840A Software Package) E8362B, E8363B, E8364B Dynamic Accuracy Test Function of the Test: To measure the relative power linearity of the analyzer’s receivers. Specification Tested: Test Port Input–Dynamic Accuracy Equipment Used: The dynamic accuracy test set (Z5623A Option H01), a power meter (E4418B or E4419B) power sensors, and two test cables.
Page 71
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Performance Tests (Agilent N7840A Software Package) If the Analyzer Fails this Test: • A failure of this test indicates a failure of one or more of the following assemblies.
Tests and Adjustments PNA Series Microwave Network Analyzers Adjustments E8362B, E8363B, E8364B Adjustments These adjustments are firmware-driven tests that are used to fine-tune your analyzer for optimum response. If multiple adjustments are to be performed, perform them in the order listed.
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Adjustments 10 MHz Frequency Reference Adjustment The 10 MHz frequency adjustment is used to adjust the frequency accuracy of the network analyzer’s 10 MHz frequency reference on the A10 frequency reference board assembly.
Tests and Adjustments PNA Series Microwave Network Analyzers Adjustments E8362B, E8363B, E8364B 3.8 GHz PMYO Frequency Adjustment The 3.8 GHz adjustment is used to adjust the frequency accuracy of the network analyzer’s 3.8 GHz permanent-magnet YIG oscillator (PMYO) on the A12 source 20.
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Adjustments LO Power Adjustment The LO power adjustment is used to adjust the power level of the LO signals from the A20 LO distribution assembly. This adjustment is done for only one LO output. The differences between each LO output is negligible;...
Page 76
Tests and Adjustments PNA Series Microwave Network Analyzers Adjustments E8362B, E8363B, E8364B 8. Reconnect the power and USB cables. Turn on the analyzer. 9. On the System menu, point to Service, Adjustments, and then click LO Power Adjust. 10. Ensure the GPIB settings are correct.
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Adjustments LO Offset Adjustment The LO power adjustment is used to adjust the power level of the LO signals from the A20 LO distribution assembly. This adjustment is done for only one LO output. The differences between each LO output is negligible;...
Page 78
Tests and Adjustments PNA Series Microwave Network Analyzers Adjustments E8362B, E8363B, E8364B 6. Connect the power sensor 8487A and the 11901D adapter or the E4413A power sensor to the end of the right angle adapter as shown in Figure 3-19 on page 3-43.
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Adjustments Source Calibration Adjustment The source calibration is used to adjust your network analyzer for a flat source power across its full frequency range. The differences between Port 1 and Port 2 are negligible;...
Tests and Adjustments PNA Series Microwave Network Analyzers Adjustments E8362B, E8363B, E8364B Phase Lock IF Gain Adjustment The phase lock IF gain adjustment is used to adjust the phase-lock loop IF signal amplitude to a constant level across the analyzer’s full frequency range: 1.
PNA Series Microwave Network Analyzers Tests and Adjustments E8362B, E8363B, E8364B Adjustments Receiver Calibration Adjustment The receiver calibration is used to adjust the network analyzer receivers for a flat response across its full frequency range: 1. A power meter/sensor is connected to Port 1, as shown in...
Page 82
Tests and Adjustments PNA Series Microwave Network Analyzers Adjustments E8362B, E8363B, E8364B Figure 3-22 Setup 1 for the Receiver Calibration Adjustment Figure 3-23 Setup 2 for the Receiver Calibration Adjustment Figure 3-24 Setup 3 for the Receiver Calibration Adjustment 3-50...
Troubleshooting PNA Series Microwave Network Analyzers Information in This Chapter E8362B, E8363B, E8364B Information in This Chapter The information in this chapter helps you: • Identify the portion of the analyzer at fault. • Locate the specific troubleshooting procedure to identify the assembly or peripheral at fault.
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Protect Against Electrostatic Discharge (ESD) Protect Against Electrostatic Discharge (ESD) This is important. If not properly protected against, electrostatic discharge can seriously damage your analyzer, resulting in costly repair. CAUTION To reduce the chance of electrostatic discharge, follow all of the recommendations outlined in “Electrostatic Discharge Protection”...
Check the Basics A problem can often be solved by repeating the procedure you were following when the problem occurred. Before calling Agilent Technologies or returning the instrument for service, please perform the following checks: 1. Is there power at the mains receptacle? If not, correct this situation and proceed.
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Getting Started with Troubleshooting Troubleshooting Organization Follow the flowgraph in Figure 4-1 to help direct you to the correct section for troubleshooting the analyzer. Figure 4-1 Troubleshooting Organization Flowchart Go to “Power Up Troubleshooting”...
Troubleshooting PNA Series Microwave Network Analyzers Power Up Troubleshooting E8362B, E8363B, E8364B Power Up Troubleshooting WARNING Immediately unplug the instrument from the ac power line if the unit shows any of the following symptoms: • Smoke, arcing, or unusual noise from inside the analyzer.
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Power Up Troubleshooting Power Supply Check There are no fuses to replace within the power supply. If you determine that NOTE the power supply is the failed assembly, replace the power supply.
Page 90
Troubleshooting PNA Series Microwave Network Analyzers Power Up Troubleshooting E8362B, E8363B, E8364B Measure the Individual Voltage Supplies The instrument contains potentially hazardous voltages. Refer to the WARNING safety symbols provided on the instrument and in “General Safety Considerations” on page 1-3 before operating the unit with the cover removed.
Page 91
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Power Up Troubleshooting Figure 4-3 E8356-60021 Synthesizer/Reference Extender Measurement Points Table 4-1 Extender Board Measurement Points Measurement Expected Location Signal Description Test Equipment Level (Vdc) Used power supply +5.0 −15.0 power supply −5.0...
Page 92
Troubleshooting PNA Series Microwave Network Analyzers Power Up Troubleshooting E8362B, E8363B, E8364B If All Supply Voltages Are Missing WARNING Disconnect the line-power cord before removing any assembly. Procedures described in this document may be performed with power supplied to the product while protective covers are removed.
Page 93
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Power Up Troubleshooting Isolate the remaining three assemblies: • remove the A15 CPU board • remove the A4 power supply • This leaves only the A14 system motherboard installed. If the measurements are still incorrect, this is the suspected faulty assembly.
Page 94
Troubleshooting PNA Series Microwave Network Analyzers Power Up Troubleshooting E8362B, E8363B, E8364B If the Fans Are Not Operating The power supply may be in thermal shutdown if the instrument has been CAUTION operating without the fans running. Allow the instrument to cool down before troubleshooting.
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Power Up Troubleshooting Troubleshooting LCD Display Problems This procedure is intended to isolate the faulty assembly when the display is dark. If the display is lit, but the color mix is faulty, refer to “A2 Display Test”...
Page 96
PNA Series Microwave Network Analyzers Power Up Troubleshooting E8362B, E8363B, E8364B 6. Connect the display ribbon cable and display lamp cable from A2 to A3. Connect the front panel interface ribbon cable coming from the analyzer to the A3 board.
Page 97
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Power Up Troubleshooting Table 4-4 Inverter Board Troubleshooting Steps Input Output (CN1) (CN2) Action Good Good Replace the display lamp. Refer to “Removing and Replacing the Display Inverter Board and the Display Lamp” on page 7-14.
Page 98
Troubleshooting PNA Series Microwave Network Analyzers Power Up Troubleshooting E8362B, E8363B, E8364B Verifying the A3 Front Panel Interface Board To verify that the HSYNC (horizontal sync), VSYNC (vertical sync), and LCD clock are functioning correctly, measure the signals listed in...
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Front Panel Troubleshooting Front Panel Troubleshooting The front panel assembly consists of the A1 keypad, A2 display, and the A3 front panel interface. The following tests verify the operation of the front panel assembly when the analyzer is in the measurement mode.
Page 100
Troubleshooting PNA Series Microwave Network Analyzers Front Panel Troubleshooting E8362B, E8363B, E8364B Checking the Front Panel Keys To check the front panel keys, push each key and compare the decimal value in the Value box under Keys to the key values in Table 4-6.
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Front Panel Troubleshooting Table 4-6 Keyboard Key Numbers Front Front Front Front Panel Panel Value Value Panel Key Value Value Panel Key Sweep Math/ Scale Macro Type Memory Checking the RPG (Front Panel Knob) To check the RPG knob, rotate the knob and check for a fluid movement of numbers in the box.
Troubleshooting PNA Series Microwave Network Analyzers Front Panel Troubleshooting E8362B, E8363B, E8364B • two or more consecutive “stuck on” pixels or three or more consecutive “dark” pixels (but no more than one set of two consecutive dark pixels) • “stuck on” or “dark” pixels less than 6.5 mm apart (excluding consecutive pixels) If any of these symptoms occur, replace the A2 display assembly.
Page 103
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Front Panel Troubleshooting Checking the Display To verify the display functionality, refer to “A2 Display Test” on page 4-19. Service Guide E8364-90026 4-21...
Troubleshooting PNA Series Microwave Network Analyzers Rear Panel Troubleshooting E8362B, E8363B, E8364B Rear Panel Troubleshooting Each rear panel connector is associated with a hardware group in the analyzer. You can use the data at these rear panel connectors to help troubleshoot these hardware groups in addition to testing the connectors.
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Rear Panel Troubleshooting Checking the SERIAL (RS-232), PARALLEL (1284-C), or VGA Port To verify the proper operation of the SERIAL, PARALLEL, or VGA port: • Connect a known good serial, parallel, or VGA peripheral device.
Page 106
Troubleshooting PNA Series Microwave Network Analyzers Rear Panel Troubleshooting E8362B, E8363B, E8364B Troubleshooting Systems with Controllers Passing the preceding test indicates that the analyzer's peripheral functions are operating normally. Therefore, if the analyzer has not been operating properly with an external controller, check the following: •...
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Rear Panel Troubleshooting LAN Troubleshooting Problems with the Local Area Network (LAN) can be difficult to solve. Software and protocol problems can make it difficult to determine whether the analyzer's hardware is working properly, or if there is a problem with the LAN or cabling.
Page 108
Troubleshooting PNA Series Microwave Network Analyzers Rear Panel Troubleshooting E8362B, E8363B, E8364B 5. The analyzer attempts four cycles of communications with the indicated LAN device. • It displays the time it took to complete each cycle. • Each cycle times-out after one second if no communication is established and the message, Request timed out, is displayed.
Page 109
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Rear Panel Troubleshooting Testing Between Two Analyzers The ability of the analyzer's LAN to function can be easily tested by connecting two analyzers together using a “crossover cable” (a short length of cable with an RJ-45 connector on each end).
Page 110
Troubleshooting PNA Series Microwave Network Analyzers Rear Panel Troubleshooting E8362B, E8363B, E8364B Table 4-7 LAN Pin Definitions and Wire Color Codes Pin Number Color Pin Number Color 1 (transmit +) White/orange White/blue 2 (transmit −) 6 (receive −) Orange Green...
Help About Network Analyzer analyzer application. Verify that the information displayed in this screen is correct for your analyzer. If any of the information is incorrect, contact Agilent Technologies. Refer to “Contacting Agilent” on page 2-13. Service Guide E8364-90026...
Page 112
To view this EEPROM header information on the network analyzer display, select System, Service, Utilities, View EEPROM Headers in the network analyzer application. If the information is incorrect for any of the pc boards, contact Agilent Technologies. Refer “Contacting Agilent” on page 2-13.
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting Verifying the A, B, R1, and R2 Traces (Standard S-Parameter Mode) There is no way to view the frequency offset receiver response (Option 080). NOTE However, some standard S-parameter receiver trace information is helpful in troubleshooting the frequency offset section of the PNA.
• the signal separation group — A22 switch splitter — A23 and A24 detectors (E8363B and E8364B only) — 30-dB fixed attenuators (E8362B only) — A25 and A26 test port couplers — A36 and A37 60-dB step attenuators (Option UNL only) —...
Page 115
Network . The following entry should be listed: Adapters Agilent Technologies DSP Driver #2 should be enabled. 2. If the icon to the left of the name is a yellow box containing an exclamation mark (!), use Windows Explorer to verify the presence of the following file: C:\WINNT\system32\drivers\spampnp.sys.
Page 116
Troubleshooting PNA Series Microwave Network Analyzers Measurement System Troubleshooting E8362B, E8363B, E8364B R1 and A Traces If both traces are affected, the problem is in the R1 phase lock loop. • If both traces are missing in all bands, the problem is in the A16 test set motherboard, the A28 R1 first converter, the A32 R1 receiver, or the A22 switch splitter.
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting Checking the Source Group Serial Bus Test Before performing tests on specific assemblies in the source group, it is recommended that you perform the serial bus test. This may help to isolate the failure to a specific assembly.
Page 118
Troubleshooting PNA Series Microwave Network Analyzers Measurement System Troubleshooting E8362B, E8363B, E8364B Figure 4-10 Serial Bus Test Dialog Box Node Descriptions, A12 Source 20 Node 111 ALC integrator Node 111 senses the output of the ALC integrator. The output of the ALC integrator is the combined voltage output of all compensation circuits in the ALC.
Page 119
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting Node Descriptions, A10 Frequency Reference Board Node 211 100 MHz Tune Node 211 senses the tune voltage that phase locks the 100 MHz VCO. The tune voltage is set by referencing the 100 MHz VCO to the internal 10 MHz OCXO or to the 10 MHz EXT REF.
Page 120
Troubleshooting PNA Series Microwave Network Analyzers Measurement System Troubleshooting E8362B, E8363B, E8364B Node Descriptions, A8 Fractional-N Synthesizer Board +5 Vdc Supply Voltage Node 411 senses the 5 Vdc supply voltage which Node 411 serves as a measurement reference. Node 412 1.5 to 3.0 GHz VCO Compensation Node 412 senses the average voltage...
Page 121
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting Node Descriptions, A16 Test Set Motherboard, SOMA 50 Node 611 Detector Level Node 611 senses the power level of the internal detector signal from the A22 switch splitter or the power level from the external detector from the rear panel.
Page 122
Troubleshooting PNA Series Microwave Network Analyzers Measurement System Troubleshooting E8362B, E8363B, E8364B Node Descriptions, A16 Test Set Motherboard, AUX 2 −10 V Reference Node 811 senses the −10 V supply voltage which serves Node 811 as a measurement reference. +1.78 V Bias Reference Node 812 senses the +1.78 V bias supply voltage Node 812 which serves as a measurement reference.
Page 123
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting Source Group Tests A defective A22 switch splitter can exhibit the same symptom as a faulty NOTE source. Therefore, if it is determined that the source is functioning properly, you are directed to “Checking the Signal Separation Group”...
Page 124
Troubleshooting PNA Series Microwave Network Analyzers Measurement System Troubleshooting E8362B, E8363B, E8364B A phase lock problem is due to either: • faulty RF signal generation (caused by the A12 source 20, A21 SOMA 50, A23 and A24 SOMA 70) • faulty LO signal generation (caused by the A8, A10, A11, A17 or A18 board)
Page 125
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting Checking the A21 Output (E8363B and E8364B Only) 1. Disconnect W2 from A21. 2. Connect a cable to A21 and measure the output power. A power level of a −7 dBm ±5 dB indicates the A21 assembly is functioning normally.
Page 126
Troubleshooting PNA Series Microwave Network Analyzers Measurement System Troubleshooting E8362B, E8363B, E8364B 4. Verify the presence of the 33.1667 MHz signal at A10J4. 5. If the signal is not present, replace the A10 frequency reference board. Refer to “Removing and Replacing the A6, A8, A9, and A10 Boards” on page 7-18.
Page 127
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting Checking the A8 Fractional-N Synthesizer Band 3 through 25 Output Perform this procedure if you observe a problem in bands 3 through 25 in all receivers. 1. Set the network analyzer to a CW frequency of 2 GHz.
Page 128
Troubleshooting PNA Series Microwave Network Analyzers Measurement System Troubleshooting E8362B, E8363B, E8364B Table 4-8 LO Path Frequencies Expected A17 Expected A18 Network Analyzer Network Analyzer Output Output Model Band Frequency Band Source Frequency Frequency Frequency of Problem (GHz) Setting (GHz)
Page 129
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting Checking the A18 MA 20 Output Perform this procedure if you observe a problem in any band and all signals from the A8 fractional-N synthesizer are present. 1. Refer to the block diagram at the back of this chapter and to “Bottom Semi-Rigid...
Troubleshooting PNA Series Microwave Network Analyzers Measurement System Troubleshooting E8362B, E8363B, E8364B Checking the Signal Separation Group Before checking the signal separation group, perform the following procedures: • “Getting Ready to Test” on page 4-41 Checking the Output Power of the A and B Signals Using a power meter, you can measure the outputs of the A and B signals from the front panel.
Page 131
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting Checking Port 2 Power (B Signal) The object of this check is to verify the power of the B signal across the entire frequency range. Perform this test if there is an observed problem only with the channel B trace. The ten second sweep is slow enough to allow you to observe the output power on the power meter as the sweep occurs.
Page 132
Troubleshooting PNA Series Microwave Network Analyzers Measurement System Troubleshooting E8362B, E8363B, E8364B For Port 2 measurements, set the network analyzer for an S measurement with a CW frequency of 1 GHz. Perform the following checks in the order presented. Checking the A22 Switch Splitter and A25 and A26 Test Port Couplers 1.
Page 133
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting Checking the A38 and A39 Bias Tees (Option UNL) 1. Locate the following semirigid cable at the appropriate bias tee: Options UNL and UNL/014 ..W55 for Port 1 or W56 for Port 2 2.
Troubleshooting PNA Series Microwave Network Analyzers Measurement System Troubleshooting E8362B, E8363B, E8364B Checking the Receiver Group Equipment Used for These Tests Models Model or Alternate Model Equipment Type Used With Part Number or Part Number Spectrum analyzer 8565E 856xE 1. Must be capable of measuring signals at 8.333 MHz and 41.667 MHz.
Page 135
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting Checking the A49 IF multiplexer Board Outputs Perform this procedure if a problem appears to be in one of the receiver channels. 1. Refer to the block diagram at the end of this chapter and to “Bottom Flexible Cables, All...
Troubleshooting PNA Series Microwave Network Analyzers Measurement System Troubleshooting E8362B, E8363B, E8364B Checking the Frequency Offset Group (Option 080) CAUTION This section troubleshoots only those circuits specifically related to the frequency offset group (option 080). Please refer to “Where to Begin Troubleshooting”...
Page 137
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting 3. Connect the spectrum analyzer to A10J11. 4. The spectrum analyzer should measure a signal at 5 MHz. 5. If no 5 MHz signal is present, replace the A10 frequency reference board. Refer to “Removing and Replacing the A6, A8, A9, and A10 Boards”...
Page 138
Troubleshooting PNA Series Microwave Network Analyzers Measurement System Troubleshooting E8362B, E8363B, E8364B 5. If this signal is not present, replace the A9 fractional-N synthesizer board. Refer to “Removing and Replacing the A6, A8, A9, and A10 Boards” on page 7-18.
Page 139
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting Checking the A13 Frequency Offset Receiver Band 0 Output Perform this procedure if you observe a problem in band 0 and all signals from the A12 source 20 output are present.
Page 140
Troubleshooting PNA Series Microwave Network Analyzers Measurement System Troubleshooting E8362B, E8363B, E8364B • The spectrum analyzer should measure a signal at 1 GHz. 5. If the measured signal is not present, replace A46 switch. Refer to “Removing and Replacing the A45 and A46 Reference Channel Transfer Switches (Option 081)” on page 7-70.
Page 141
PNA Series Microwave Network Analyzers Troubleshooting E8362B, E8363B, E8364B Measurement System Troubleshooting Service Guide E8364-90026 4-59...
Theory of Operation PNA Series Microwave Network Analyzers Information in This Chapter E8362B, E8363B, E8364B Information in This Chapter This chapter provides a general description of the operating theory of the PNA series microwave network analyzers. • Theory of operation is explained to the assembly level only.
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B Network Analyzer System Operation Network Analyzer System Operation The PNA series network analyzer generates a phase-locked incident signal from the internal synthesized source. By means of signal separation, the incident signal is divided into a reference signal and a test signal.
Page 154
Theory of Operation PNA Series Microwave Network Analyzers Network Analyzer System Operation E8362B, E8363B, E8364B Synthesized Source Group The built-in synthesized source generates a swept, stepped, or continuous wave (CW) signal in the frequency ranges as listed in Table 5-1. The source output power is leveled by an internal automatic leveling control (ALC) circuit.
Page 155
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B Network Analyzer System Operation IF Access Group This group provides the circuitry needed for the IF access (Option H11) capability. Option H11 is also necessary for the operation of the pulse measurement capability, Option H08.
Theory of Operation PNA Series Microwave Network Analyzers Source Group Operation E8362B, E8363B, E8364B Source Group Operation The source group produces a stable output signal by phase locking a yttrium-iron-garnet (YIG) oscillator to a synthesized voltage-controlled oscillator (VCO). Refer to...
Page 157
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B Source Group Operation Figure 5-2 Source Group Service Guide E8364-90026...
Theory of Operation PNA Series Microwave Network Analyzers Source Group Operation E8362B, E8363B, E8364B Table 5-2 Subsweep Frequencies A8 Frac-N A20 LODA Synthesizer LOMA 10 (L.O.) Source 20 SOMA 50 Frequency Frequency Frequency Frequency Frequency (GHz) (GHz) (GHz) (GHz) (GHz) (E8363/4B Only) 0.011 to 0.046 0.011 to 0.046 0.011 to 0.046 0.010 to 0.045 0.010 to 0.045...
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B Source Group Operation At the output, an ALC circuit maintains the power level at −5 dBm ± 2 dB for bands 0–1 or −1.5 dBm ± 2 dB for bands 2–25. In all bands, the output is sent to the A17–A20 LO distribution path and on to the A27–A30 first converters, as the 1st LO signal.
Theory of Operation PNA Series Microwave Network Analyzers Source Group Operation E8362B, E8363B, E8364B Rear-Panel Interconnects 10 MHz REF. INPUT A BNC connector that allows an external frequency reference signal to be used to phase lock the analyzer for increased frequency accuracy.
Together, these two paths produce an output frequency range of 10 MHz to 20 GHz. For the E8362B analyzer, this signal is sent directly to the A22 switch splitter in the signal separation group. For the E8363B and E8364B analyzers, this signal is sent to the A21 source multiplier/amplifier 50 (SOMA 50).
Theory of Operation PNA Series Microwave Network Analyzers Source Group Operation E8362B, E8363B, E8364B A16 Test Set Motherboard The A16 test set motherboard serves these functions: • to act as an interface between the A15 CPU board and the auxiliary rear panel interconnects.
Page 163
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B Source Group Operation Table 5-3 TEST SET I/O Connector Pin Assignments DB-25 Female Connector Pin Numbers Name Function SEL0 TTL out, test set select bit 0, tied to 0 V...
Page 164
Theory of Operation PNA Series Microwave Network Analyzers Source Group Operation E8362B, E8363B, E8364B Table 5-4 HANDLER I/O Connector Key Pin Assignments Rectangular 36-Pin Female Connector Pin Numbers Name Function 0 V, ground reference TTL in, negative pulse (1 μs min) latches OUTPUT1-2 INPUT1 3–4...
Page 165
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B Source Group Operation Table 5-5 AUX I/O Connector Pin Assignments DB-25 Male Connector Pin Numbers Name Function ACOM 0 V, ground reference for analog signals −10 to +10 Vdc output, 10 mA max, Ro = 100 Ω...
For the E8363B and E8364B, the output of the A21 SOMA 50 is routed to the A22 switch splitter. For the E8362B, this is the output of the A12 source 20 assembly. The A22 switch splitter is a solid-state switch. It switches between the Port 1 and Port 2 measurement paths, automatically enabling alternate forward and reverse measurements.
The A23 and A24 detectors sense the R1 and R2 reference signal levels and provide feedback to the ALC circuitry on the A16 test set motherboard. 30-dB Fixed Attenuators (E8362B Only) The 30-dB fixed attenuators are placed in the reference signal paths to provide isolation.
• the A25 test port 1 coupler and the A27 receiver A first converter • the A26 test port 2 coupler and the A30 receiver B first converter • the A23 detector (for E8363B and E8364B) or fixed attenuator (for E8362B) and the A28 receiver R1 first converter •...
Page 169
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B Signal Separation Group Operation Normal Option 014 Configuration The normal configuration for the Option 014 is shown in Figure 5-4 on page 5-19. With this configuration and inclusion of an external amplifier and accessories, you can calibrate the analyzer and test devices at power levels up to +30 dBm.
Page 170
Theory of Operation PNA Series Microwave Network Analyzers Signal Separation Group Operation E8362B, E8363B, E8364B High Dynamic Range Configuration With a few jumper changes, you can configure the Option 014 for higher dynamic range measurements. By swapping the front panel jumpers for one port, signal flow through the corresponding coupler is reversed, increasing the test signal sensitivity by 15 dB.
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B Receiver Group Operation Receiver Group Operation The receiver group measures and processes the input signals into digital information for processing and eventual display. Figure 5-6 on page 5-22 is a simplified block diagram of the receiver functional group.
Page 172
Theory of Operation PNA Series Microwave Network Analyzers Receiver Group Operation E8362B, E8363B, E8364B Figure 5-6 Receiver Group 5-22 Service Guide E8364-90026...
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B Receiver Group Operation A6 SPAM Board (Analog Description) The A6 SPAM board contains digital and analog circuitry. For digital descriptions, refer to “A6 SPAM Board (Digital Description)” on page 5-32.
Theory of Operation PNA Series Microwave Network Analyzers Frequency Offset Group Operation (Options 080 and 081) E8362B, E8363B, E8364B Frequency Offset Group Operation (Options 080 and 081) The frequency offset group provides the circuitry necessary to phase lock the analyzer while the source and receiver are tuned to different frequencies.
Page 175
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B Frequency Offset Group Operation (Options 080 and 081) Figure 5-7 Frequency Offset Group Service Guide E8364-90026 5-25...
The A47 RF path coupler is located in the RF path of the PNA between the A12 source and either the A21 SOMA 50 (for E8363/4B) or the A22 switch (for E8362B). A portion of the RF signal is coupled off and delivered to the rear-panel TEST SET DRIVERS RF connector.
Page 177
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B IF Access Group (Option H11) A49 IF multiplexer board where they can be switched into the signal path of each of the four measurement channels. Service Guide E8364-90026 5-27...
Page 178
Theory of Operation PNA Series Microwave Network Analyzers IF Access Group (Option H11) E8362B, E8363B, E8364B Figure 5-8 IF Access Group 5-28 Service Guide E8364-90026...
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B Digital Processing and Digital Control Group Operation Digital Processing and Digital Control Group Operation The digital processor and control group provides digital control for the entire analyzer. It provides: •...
Page 180
Theory of Operation PNA Series Microwave Network Analyzers Digital Processing and Digital Control Group Operation E8362B, E8363B, E8364B Figure 5-9 Digital Processing and Digital Control Group 5-30 Service Guide E8364-90026...
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B Digital Processing and Digital Control Group Operation Front Panel Subgroup The front panel subgroup contains the following assemblies: • A1 Keypad Assembly • A2 Display Assembly • A3 Front Panel Interface Board A1 Keypad Assembly The A1 keypad assembly provides user interface to the analyzer.
Theory of Operation PNA Series Microwave Network Analyzers Digital Processing and Digital Control Group Operation E8362B, E8363B, E8364B Data Acquisition and Processing Subgroup The data acquisition and processing subgroup contain the following assemblies. See Figure 5-9 on page 5-30. •...
Page 183
PNA Series Microwave Network Analyzers Theory of Operation E8362B, E8363B, E8364B Digital Processing and Digital Control Group Operation Rear Panel Interconnects The rear panel includes the following interfaces: A universal serial bus (USB) jack (an industry standard 4-pin connector). GPIB A 24-pin, female, type D-24 connector that meets IEEE-488 standards.
Theory of Operation PNA Series Microwave Network Analyzers Power Supply Group Operation E8362B, E8363B, E8364B Power Supply Group Operation The A4 power supply assembly is a switching power supply operating at 103 kHz switching frequency. The input power ranges for the power supply are 90 to 132 Vac or 195 to 250 Vac.
The definition of a rebuilt-exchange assembly. Rebuilt-Exchange Assemblies Page 6-5 The procedure for replacing and returning a defective assembly to Agilent Technologies. This section lists the assemblies by their reference number and illustrates the location of the replaceable parts in the analyzer, including: •...
To order a part listed in the replaceable parts lists: • include the part number • indicate the quantity required • Contact Agilent Technologies for instructions on where to send the order. Refer to “Contacting Agilent” on page 2-13. To order a part that is not listed in the replaceable parts lists: •...
Replaceable Parts PNA Series Microwave Network Analyzers Assembly Replacement Sequence E8362B, E8363B, E8364B Assembly Replacement Sequence The following steps show the sequence that you should follow to replace an assembly in the network analyzer. Step 1. Identify the faulty group. Begin with Chapter 4, “Troubleshooting.”...
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Rebuilt-Exchange Assemblies Rebuilt-Exchange Assemblies Under the rebuilt-exchange assembly program: • Certain factory-repaired and tested assemblies are available on a trade-in basis. • Exchange assemblies are offered for lower cost than a new assembly, but meet all factory specifications required of a new assembly.
Splitter Local oscillator distribution assembly (LODA) “Bottom Assemblies, No Options” on page 6-20 Source multiplier/amplifier 50 (SOMA 50) (E8363B, E8364B only) Switch splitter None Channel R1 30-dB attenuator (E8362B only) None Channel R2 30-dB attenuator (E8362B only) Service Guide E8364-90026...
Page 191
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Table 6-1 Location by Reference Designator (Continued) Reference Designator Description Location Channel R1 detector (E8363B, E8364B only) Channel R2 detector (E8363B, E8364B only) Test port 1 coupler...
Page 192
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Table 6-2 Part Number Location by Type of Part Type of Part Location • “Front Panel Assembly, All Options” on page 6-10 • “Top Assemblies, All Options Except 080” on page 6-12 •...
Page 193
• GPIB Cables/GPIB Adapter on page 6-80 Miscellaneous • ESD Supplies on page 6-83 • E8362B Upgrade Kits on page 6-81 • E8363B, E8364B Upgrade Kits on page 6-81 • Touch-up Paint on page 6-82 • USB Accessories on page 6-82 •...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Front Panel Assembly, All Options Ref. Desig. Model Part Number Qty Description E8356-60001 Keypad assembly E8356-60002 Display assembly 0950-3379 Inverter board (included in A2 2090-0386 Display lamp for LCD (included in A2...
Page 195
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-2 Front Panel Assembly, All Options Service Guide E8364-90026 6-11...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Top Assemblies, All Options Except 080 Ref. Desig. Part Number Description 0950-4599 Power supply assembly E8356-60006 Signal processing ADC module (SPAM) board E8364-60099 Fractional-N synthesizer board E8364-60136...
Page 197
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-3 Top Assemblies, All Options Except 080 Service Guide E8364-90026 6-13...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Top Cables–All, All Options Except 080 Ref. Desig. Type Part Number Description A8 fractional-N synthesizer board J106 E8364-20027 to A17 LOMA 10 J2 A8 fractional-N synthesizer board J101...
Page 199
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-4 Top Cables–All, All Options Except 080 Service Guide E8364-90026 6-15...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Top Assemblies, All Options Including 080 Ref. Desig; Part Number Description 0950-4599 Power supply assembly E8356-60006 Signal processing ADC module (SPAM) board E8364-60099 Fractional-N synthesizer board E8364-60099...
Page 201
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-5 Top Assemblies, All Options Including 080 Service Guide E8364-90026 6-17...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Top Cables–All, All Options Including 080 Ref. Desig. Type Part Number Description A8 fractional-N synthesizer board J106 E8364-20027 to A17 LOMA 10 J2 A8 fractional-N synthesizer board J101...
Page 203
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-6 Top Cables–All, All Options Including 080 Service Guide E8364-90026 6-19...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Assemblies, No Options Ref. Desig. Models Part Number Qty Description E8364-60138 Test set motherboard 5087-7127 LOMA 10 5087-6127 5087-7128 MA 20 5087-6128 5087-7139 Splitter 5087-7129 LODA...
Page 205
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-7 Bottom Assemblies, No Options Service Guide E8364-90026 6-21...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Assemblies, Option UNL Ref. Desig. Models Part Number Qty Description E8364-60138 Test set motherboard 5087-7127 LOMA 10 5087-6127 5087-7128 MA 20 5087-6128 5087-7139 Splitter 5087-7129 LODA...
Page 209
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-9 Bottom Assemblies, Option UNL Service Guide E8364-90026 6-25...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Assemblies, Option 014 and Options 014/080 Ref. Desig. Models Part Number Qty Description E8364-60138 Test set motherboard 5087-7127 LOMA 10 5087-6127 5087-7128 MA 20 5087-6128 5087-7139...
Page 213
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-11 Bottom Assemblies, Option 014 and Options 014/080 Service Guide E8364-90026 6-29...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Assemblies, Options UNL/014 and UNL/014/080 Ref. Desig. Models Part Number Qty Description E8364-60138 Test set motherboard 5087-7127 LOMA 10 5087-6127 5087-7128 MA 20 5087-6128 5087-7139 Splitter...
Page 217
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-13 Bottom Assemblies, Options UNL/014 and UNL/014/080 Service Guide E8364-90026 6-33...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Assemblies, Options UNL/016 Ref. Desig. Models Part Number Qty Description E8364-60138 Test set motherboard 5087-7127 LOMA 10 5087-6127 5087-7128 MA 20 5087-6128 5087-7139 Splitter 5087-7129 LODA...
Page 221
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-15 Bottom Assemblies, Options UNL/016 Service Guide E8364-90026 6-37...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Assemblies, Options UNL/014/016 and UNL/014/016/080 Ref. Desig. Models Part Number Qty Description E8364-60138 Test set motherboard 5087-7127 LOMA 10 5087-6127 5087-7128 MA 20 5087-6128 5087-7139 Splitter...
Page 225
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-17 Bottom Assemblies, Options UNL/014/016 and UNL/014/016/080 Service Guide E8364-90026 6-41...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Assemblies, Options 014/080/081 Ref. Desig. Models Part Number Qty Description E8364-60138 Test set motherboard 5087-7127 LOMA 10 5087-6127 5087-7128 MA 20 5087-6128 5087-7139 Splitter 5087-7129 LODA...
Page 229
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-19 Bottom Assemblies, Options 014/080/081 Service Guide E8364-90026 6-45...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Assemblies, Options UNL/014/080/081 Ref. Desig. Models Part Number Qty Description E8364-60138 Test set motherboard 5087-7127 LOMA 10 5087-6127 5087-7128 MA 20 5087-6128 5087-7139 Splitter 5087-7129 LODA...
Page 233
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-21 Bottom Assemblies, Options UNL/014/080/081 Service Guide E8364-90026 6-49...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Assemblies, Options UNL/014/080/081/H11 Ref. Desig. Models Part Number Qty Description E8364-60138 Test set motherboard 5087-7127 LOMA 10 5087-6127 5087-7128 MA 20 5087-6128 5087-7139 Splitter 5087-7129 LODA...
Page 237
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-23 Bottom Assemblies, Options UNL/014/080/081/H11 Service Guide E8364-90026 6-53...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Assemblies, Options UNL/014/016/080/081 Ref. Desig. Models Part Number Qty Description E8364-60138 Test set motherboard 5087-7127 LOMA 10 5087-6127 5087-7128 MA 20 5087-6128 5087-7139 Splitter 5087-7129 LODA...
Page 241
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-25 Bottom Assemblies, Options UNL/014/016/080/081 Service Guide E8364-90026 6-57...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Assemblies, Options UNL/014/016/080/081/H11 Ref. Desig. Models Part Number Qty Description E8364-60138 Test set motherboard 5087-7127 LOMA 10 5087-6127 5087-7128 MA 20 5087-6128 5087-7139 Splitter 5087-7129 LODA...
Page 245
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-27 Bottom Assemblies, Options UNL/014/016/080/081/H11 Service Guide E8364-90026 6-61...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Flexible Cables, All Options Ref. Desig. Models Part Number Qty Description A27 channel A mixer E8364-60019 to A49 IF multiplexer board A28 channel R1 mixer E8364-60020...
Page 249
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-29 Bottom Flexible Cables, All Options Service Guide E8364-90026 6-65...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Ribbon Cables and Wrapped-Wire, All Options Ref. Desig. Type Part Number Description A41 hard disk drive assembly ➀ 8121-0817 to A15 CPU board A16 test set motherboard P701 (MAT HDLR) ➁...
Page 251
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-30 Bottom Ribbon Cables and Wrapped Wire, All Options Service Guide E8364-90026 6-67...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Top Hardware, All Options Ref. Desig. Part Number Description Fans Refer to “Internal Hardware and Parts, All Options” on page 6-72. ➀ 5022-6134 Cable hold down ➁...
Page 253
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-31 Top Hardware, All Options Service Guide E8364-90026 6-69...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Bottom Hardware, All Options Ref. Desig. Part Number Description ➀ E8364-00002 Detector/attenuator/bias tee bracket Machine screw M3.0 x 6 pan head ➁ 0515-0430 (Detector brackets-3 ea; front-panel LED board-3; A19 splitter bracket-2;...
Page 255
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-32 Bottom Hardware, All Options Service Guide E8364-90026 6-71...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Internal Hardware and Parts, All Options Ref. Desig. Models Part Number Description 0950-2782 Floppy disk drive 8121-0120 Cable E4406-60014 Floppy disk drive board E4406-40006 Floppy disk drive mount...
Page 257
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-33 Internal Hardware and Parts, All Options Service Guide E8364-90026 6-73...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B External Hardware and Parts, All Options Ref. Desig. Part Number Description Keypad assembly Display assembly Refer to “Front Panel Assembly, All Options” on page 6-10. Front knob ➀...
Page 259
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-34 External Hardware and Parts, All Options Service Guide E8364-90026 6-75...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Hard Disk Drive Assembly Reference Designator Part Number Description Hard disk drive assembly kit with N8980A Kit (for 500 MHz CPU board) programmed hard disk drive (HDD) ➀...
Page 263
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Figure 6-36. Hard Disk Drive Assembly Service Guide E8364-90026 6-79...
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Miscellaneous Part Numbers Table 6-3 Part Numbers for Miscellaneous Parts and Accessories Model or Part Description Number Service Tools Extender board, synthesizer/reference E8356-60021 20-mm open-end torque wrench (included with analyzer); 0.9 N-m (8 in-lb)
Page 265
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Table 6-3 Part Numbers for Miscellaneous Parts and Accessories (Continued) Model or Part Description Number E8362B Upgrade Kits Time domain (Option 010) E8362BU-010 Configurable test set (Option 014)
Page 266
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B Table 6-3 Part Numbers for Miscellaneous Parts and Accessories (Continued) Model or Part Description Number Add built-in performance test software for standards compliant calibration perpetual license E8363/4BU-898...
Page 267
PNA Series Microwave Network Analyzers Replaceable Parts E8362B, E8363B, E8364B Replaceable Parts Listings Table 6-3 Part Numbers for Miscellaneous Parts and Accessories (Continued) Model or Part Description Number ESD Supplies Adjustable antistatic wrist strap 9300-1367 Antistatic wrist strap grounding cord (5 foot length)
Page 268
Replaceable Parts PNA Series Microwave Network Analyzers Replaceable Parts Listings E8362B, E8363B, E8364B 6-84 Service Guide E8364-90026...
Information in This Chapter E8362B, E8363B, E8364B Information in This Chapter This chapter contains procedures for removing and replacing the major assemblies of your Agilent Technologies PNA series microwave network analyzer. Chapter Seven at-a-Glance Section Title Summary of Content Start Page...
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Before Starting the Removal and Replacement Procedures Before Starting the Removal and Replacement Procedures Before starting the removal and replacement procedures, review these warnings, cautions, and the assembly replacement sequence.
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Raising and Removing the Receiver Deck E8362B, E8363B, E8364B Raising and Removing the Receiver Deck Tools Required • T-10 TORX driver (set to 9 in-lb) Raising the Receiver Deck Refer to Figure 7-2 for this procedure.
Page 277
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Raising and Removing the Receiver Deck Figure 7-2 Receiver Deck, Raising and Removal Service Guide E8364-90026...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the Front Panel Assembly E8362B, E8363B, E8364B Removing and Replacing the Front Panel Assembly Tools Required • T-10 TORX driver (set to 9 in-lb) • T-20 TORX driver (set to 21 in-lb) •...
Page 279
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the Front Panel Assembly Figure 7-3 Front Panel Assembly Removal Service Guide E8364-90026 7-11...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing Front Panel Subassemblies E8362B, E8363B, E8364B Removing and Replacing Front Panel Subassemblies Tools Required • T-10 TORX driver (set to 9 in-lb) • T-20 TORX driver (set to 21 in-lb) •...
Page 281
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing Front Panel Subassemblies Replacement Procedure When replacing the assemblies, ensure that the ribbon cables are not under NOTE the assemblies. 1. Reverse the order of the removal procedure.
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the Display Inverter Board and the Display Lamp E8362B, E8363B, E8364B Removing and Replacing the Display Inverter Board and the Display Lamp Tools Required • T-10 TORX driver (set to 9 in-lb) •...
Page 283
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the Display Inverter Board and the Display Lamp NOTE To avoid dust or dirt particles from getting in between the display glass and the display, do not completely take out the display.
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A4 Power Supply Assembly E8362B, E8363B, E8364B Removing and Replacing the A4 Power Supply Assembly Tools Required • T-10 TORX driver (set to 9 in-lb) • T-20 TORX driver (set to 21 in-lb) •...
Page 285
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A4 Power Supply Assembly Figure 7-6 A4 Power Supply Assembly Removal Service Guide E8364-90026 7-17...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A6, A8, A9, and A10 Boards E8362B, E8363B, E8364B Removing and Replacing the A6, A8, A9, and A10 Boards Tools Required • T-10 TORX driver (set to 9 in-lb) •...
Page 287
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A6, A8, A9, and A10 Boards Figure 7-7 A6, A8, A9, and A10 Board Assemblies Removal Service Guide E8364-90026 7-19...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A13 Frequency Offset Receiver Assembly E8362B, E8363B, E8364B Removing and Replacing the A13 Frequency Offset Receiver Assembly Tools Required • T-10 TORX driver (set to 9 in-lb) •...
Page 293
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A13 Frequency Offset Receiver Assembly Figure 7-10 A13 Frequency Offset Receiver Assembly Removal Service Guide E8364-90026 7-25...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A14 System Motherboard E8362B, E8363B, E8364B Removing and Replacing the A14 System Motherboard Tools Required • T-10 TORX driver (set to 9 in-lb) • T-20 TORX driver (set to 21 in-lb) •...
Page 295
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A14 System Motherboard Figure 7-11 A14 System Motherboard Removal Service Guide E8364-90026 7-27...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A15 CPU Board E8362B, E8363B, E8364B Removing and Replacing the A15 CPU Board Tools Required • T-10 TORX driver (set to 9 in-lb) • T-20 TORX driver (set to 21 in-lb) •...
Page 297
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A15 CPU Board Figure 7-12 A15 CPU Board Removal Service Guide E8364-90026 7-29...
“Raising and Removing the Receiver Deck” on page 7-8 if necessary. 3. Remove the A21 SOMA 50 (E8363B and E8364B) or RF cable W1 (E8362B). Refer to “Removing and Replacing the A21 SOMA 50 (E8363B and E8364B Only)” on page 7-40.
Page 299
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A16 Test Set Motherboard Figure 7-13 A16 Test Set Motherboard Removal Service Guide E8364-90026 7-31...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A17 LOMA 10 E8362B, E8363B, E8364B Removing and Replacing the A17 LOMA 10 Tools Required • T-10 TORX driver (set to 9 in-lb) • T-20 TORX driver (set to 21 in-lb) •...
Page 301
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A17 LOMA 10 Figure 7-14 A17 LOMA 10 Removal Service Guide E8364-90026 7-33...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A18 MA 20 E8362B, E8363B, E8364B Removing and Replacing the A18 MA 20 Tools Required • T-10 TORX driver (set to 9 in-lb) • T-20 TORX driver (set to 21 in-lb) •...
Page 303
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A18 MA 20 Figure 7-15 A18 MA 20 Removal Service Guide E8364-90026 7-35...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A21 SOMA 50 (E8363B and E8364B Only) E8362B, E8363B, E8364B Removing and Replacing the A21 SOMA 50 (E8363B and E8364B Only) Tools Required • T-10 TORX driver (set to 9 in-lb) •...
Page 309
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A21 SOMA 50 (E8363B and E8364B Only) Figure 7-18 A21 SOMA 50 Removal (E8363B and E8364B Only) Service Guide E8364-90026 7-41...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A23 and A24 Detectors (E8363B and E8364B only) E8362B, E8363B, E8364B Removing and Replacing the A23 and A24 Detectors (E8363B and E8364B only) Tools Required • T-10 TORX driver (set to 9 in-lb) •...
Page 315
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A23 and A24 Detectors (E8363B and E8364B only) Figure 7-21 A23 and A24 Detectors Removal (E8363B and E8364B Only) Service Guide E8364-90026 7-47...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A25 and A26 Test Port Couplers E8362B, E8363B, E8364B Removing and Replacing the A25 and A26 Test Port Couplers Tools Required • T-10 TORX driver (set to 9 in-lb) •...
Page 317
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A25 and A26 Test Port Couplers Figure 7-22 A25 and A26 Test Port Couplers Removal Service Guide E8364-90026 7-49...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A27, A28, A29, and A30 First Converters (Mixers) and Mixer Bias Board E8362B, E8363B, E8364B Removing and Replacing the A27, A28, A29, and A30 First Converters (Mixers) and Mixer Bias Board Tools Required •...
Page 319
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A27, A28, A29, and A30 First Converters (Mixers) and Mixer Bias Board Figure 7-23 A27, A28, A29, and A30 First Converters (Mixers) Removal Service Guide E8364-90026...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A31, A32, A33, and A34 Receiver Modules and A49 IF Multiplexer Board E8362B, E8363B, E8364B Removing and Replacing the A31, A32, A33, and A34 Receiver Modules and A49 IF Multiplexer Board Tools Required •...
Page 321
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A31, A32, A33, and A34 Receiver Modules and A49 IF Multiplexer Board Figure 7-24 A31, A32, A33, and A34 Modules and A49 IF Multiplexer Board...
Receiver Deck” on page 7-8 if necessary. 3. Identify the step attenuator to be removed. To replace the attenuator, the bracket holding the attenuator, detector (E8363B/E8364B) or fixed pad (E8362B), and bias tee must be removed. ① 4. Disconnect the ribbon cable (item ) from the appropriate attenuator.
Page 325
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A36 and A37 60-dB Source Step Attenuators (Option UNL) Figure 7-26 A36 and A37 60-dB Source Step Attenuators Removal (Option UNL) Service Guide E8364-90026...
Receiver Deck” on page 7-8 if necessary. 3. Identify the bias tee to be removed. To replace the bias tee, the bracket holding the attenuator, detector (E8363B/E8364B) or fixed pad (E8362B), and bias tee must be removed. ① 4. Disconnect the ribbon cable (item ) from the appropriate attenuator.
Page 327
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A38 and A39 Bias Tees (Option UNL) Figure 7-27 A38 and A39 Bias Tees Removal (Option UNL) Service Guide E8364-90026 7-59...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A40 Floppy Disk Drive E8362B, E8363B, E8364B Removing and Replacing the A40 Floppy Disk Drive Tools Required • T-10 TORX driver (set to 9 in-lb) • T-20 TORX driver (set to 21 in-lb) •...
Page 329
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A40 Floppy Disk Drive Figure 7-28 A40 Floppy Disk Drive Removal Figure 7-29 A40 Floppy Disk Drive (Exploded View) Service Guide E8364-90026 7-61...
PNA Series Microwave Network Analyzers Removing and Replacing the A41 Hard Disk Drive (HDD) E8362B, E8363B, E8364B Removing and Replacing the A41 Hard Disk Drive (HDD) Certain unique files exist on the hard disk drive that are necessary for proper operation of your analyzer.
Page 331
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A41 Hard Disk Drive (HDD) Figure 7-30 Hard Disk Drive Assembly Removal Disassembling the HDDA If the HDD is being replaced by the HDDA (N8980A or N8981A), it is not NOTE necessary to perform these steps.
Page 332
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A41 Hard Disk Drive (HDD) E8362B, E8363B, E8364B Figure 7-31 Hard Disk Drive Assembly (Exploded View) 7-64 Service Guide E8364-90026...
Page 333
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A41 Hard Disk Drive (HDD) Reassembly Procedure NOTE If the HDD is being replaced by the HDDA (N8980A or N8981A), it is not necessary to perform these steps.
Page 334
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A41 Hard Disk Drive (HDD) E8362B, E8363B, E8364B Initialize the Windows Operating System 1. Connect a mouse and keyboard to the analyzer. 2. Turn on power to the analyzer.
Page 335
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A41 Hard Disk Drive (HDD) 3. Point to Programs, Accessories, and then open Windows Explorer. 4. Navigate to C:\Program Files\Agilent\Network Analyzer. 5. Copy each of the backup files from the floppy disk or USB memory drive to the HDD.
Page 336
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A43 and A44 35-dB Receiver Step Attenuators (Option 016) E8362B, E8363B, E8364B Removing and Replacing the A43 and A44 35-dB Receiver Step Attenuators (Option 016) Tools Required •...
Page 337
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A43 and A44 35-dB Receiver Step Attenuators (Option 016) Figure 7-32 A43 and A44 35-dB Receiver Step Attenuators Removal (Option 016) Service Guide E8364-90026...
Page 338
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A45 and A46 Reference Channel Transfer Switches (Option 081) E8362B, E8363B, E8364B Removing and Replacing the A45 and A46 Reference Channel Transfer Switches (Option 081) Tools Required •...
Page 339
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A45 and A46 Reference Channel Transfer Switches (Option 081) Figure 7-33 A45 and A46 Reference Channel Transfer Switches Removal (Option 081) Service Guide E8364-90026...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the A47 and A48 Couplers (Option H11) E8362B, E8363B, E8364B Removing and Replacing the A47 and A48 Couplers (Option H11) Tools Required • T-10 TORX driver (set to 9 in-lb) •...
Page 341
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the A47 and A48 Couplers (Option H11) Figure 7-34 A47 and A48 Couplers Removal (Option H11) Service Guide E8364-90026 7-73...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the Midweb and the B1 Fan E8362B, E8363B, E8364B Removing and Replacing the Midweb and the B1 Fan Tools Required • T-10 TORX driver (set to 9 in-lb) •...
Page 343
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the Midweb and the B1 Fan Removing the Midweb Refer to Figure 7-35. ① 1. Disconnect the power cable (item ) for the three fans from the A14 system motherboard.
Page 344
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Removing and Replacing the Midweb and the B1 Fan E8362B, E8363B, E8364B Removing the Fan Refer to Figure 7-36. 1. To remove the fan, it is necessary to remove the four nylon rivets that attach it to the ➀...
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Removing and Replacing the USB Hub Removing and Replacing the USB Hub Tools Required • T-10 TORX driver (set to 9 in-lb) • ESD grounding wrist strap Removal Procedure...
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Post-Repair Procedures E8362B, E8363B, E8364B Post-Repair Procedures After the replacement of an assembly, you must perform the service procedures in the order listed in Table 7-2. Procedures referenced in this table are located in Chapter 3, “Tests and...
Page 349
PNA Series Microwave Network Analyzers Repair and Replacement Procedures E8362B, E8363B, E8364B Post-Repair Procedures Table 7-2 Related Service Procedures (Continued) Adjustments Verification, Performance, Replaced Assembly and Other Procedures and Other Tests and Procedures Reinstall the serial number. (Refer to “Installing or Changing a Serial Number”...
Page 350
Repair and Replacement Procedures PNA Series Microwave Network Analyzers Post-Repair Procedures E8362B, E8363B, E8364B Table 7-2 Related Service Procedures (Continued) Adjustments Verification, Performance, Replaced Assembly and Other Procedures and Other Tests and Procedures A35 receiver Receiver Calibration The Operator’s Check...
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers Information in This Chapter E8362B, E8363B, E8364B Information in This Chapter Chapter Eight at-a-Glance Section Title Summary of Content Start Page Error Terms How to use error terms as a preventive page 8-3 maintenance and troubleshooting tool.
PNA Series Microwave Network Analyzers General Purpose Maintenance Procedures E8362B, E8363B, E8364B Error Terms Error Terms Using Error Terms as a Diagnostic Tool By examining error terms, you can monitor system performance for preventive maintenance and troubleshooting purposes. The magnitude and shape of the error terms are affected by: •...
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers Error Terms E8362B, E8363B, E8364B Performing Measurement Calibration A calibration must be performed to allow the analyzer to calculate the error terms before they can be used as a tool: Perform the following procedure only at a static-safe workstation, and wear a CAUTION grounded wrist strap.
Page 355
PNA Series Microwave Network Analyzers General Purpose Maintenance Procedures E8362B, E8363B, E8364B Error Terms Figure 8-2 Flowgraph of Two-Port Error Terms where: E = error term 1st Subscript: D = Directivity S = Source Match R = Reflection Tracking X = Crosstalk (Isolation)
Page 356
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers Error Terms E8362B, E8363B, E8364B Figure 8-3 Flowgraph of Three-Port Error Terms where: E = error term DIR = Directivity MAT = Forward Source Match and Reverse Load Match TRK = Forward Reflection Tracking and Reverse Transmission Tracking...
Page 357
PNA Series Microwave Network Analyzers General Purpose Maintenance Procedures E8362B, E8363B, E8364B Error Terms Figure 8-4 Flowgraph of Four-Port Error Terms where: E = error term DIR = Directivity MAT = Forward Source Match and Reverse Load Match TRK = Forward Reflection Tracking and Reverse Transmission Tracking...
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers Error Terms E8362B, E8363B, E8364B Accessing Error Terms Error terms can be accessed either manually or programmatically: Manually • “Front Panel Access to Error Terms” on page 8-8 Programmatically • “GPIB Access to Error Terms” on page 8-9 •...
Page 359
PNA Series Microwave Network Analyzers General Purpose Maintenance Procedures E8362B, E8363B, E8364B Error Terms Programmatic Access to Error Terms GPIB Access to Error Terms You can access error terms by way of GPIB with Standard Commands for Programmable Instruments (SCPI).
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers Error Terms E8362B, E8363B, E8364B Error Term Data The error term descriptions in this section include the following information: • a table of the error terms • description and significance of each error term •...
Page 361
PNA Series Microwave Network Analyzers General Purpose Maintenance Procedures E8362B, E8363B, E8364B Error Terms Data are listed here as a convenience only. Detailed instrument specifications IMPORTANT are listed in the embedded help in the network analyzer. Table 8-1 Error Term Data...
Page 362
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers Error Terms E8362B, E8363B, E8364B Directivity (E and E and E are the uncorrected forward and reverse directivity error terms of the system. The directivity error of the test port is determined by measuring the S and S reflection of the calibration kit load.
Page 363
PNA Series Microwave Network Analyzers General Purpose Maintenance Procedures E8362B, E8363B, E8364B Error Terms Reflection Tracking (E and E Reflection tracking is the difference between the frequency response of the reference path (R1 or R2path) and the frequency response of the reflection test path (A or B input path).
Page 364
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers Error Terms E8362B, E8363B, E8364B Isolation (E and E Isolation, or crosstalk, is the uncorrected forward and reverse isolation error terms that represent leakage between the test ports and the signal paths. The isolation error terms...
PNA Series Microwave Network Analyzers General Purpose Maintenance Procedures E8362B, E8363B, E8364B Option Enable Utility Option Enable Utility Accessing the Option Enable Utility To start the option enable utility: • On the menu, point to , and then click System...
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers Option Enable Utility E8362B, E8363B, E8364B Enabling or Removing Options There are two types of options: • Hardware: Hardware options involve adding additional hardware to the analyzer. After the proper hardware has been installed in the analyzer, the option can be enabled using the option enable utility.
PNA Series Microwave Network Analyzers General Purpose Maintenance Procedures E8362B, E8363B, E8364B Option Enable Utility Repairing and Recovering Option Data License, option, and model number data are stored in an EEPROM on the frequency reference board and written into the gen.lic file on the hard disk drive. If the data stored in either of these items is lost, it can be recovered from the other item.
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers Option Enable Utility E8362B, E8363B, E8364B Installing or Changing a Serial Number It is necessary to reinstall the instrument serial number if the test set motherboard is replaced. IMPORTANT Use extreme care when entering the serial number, as only one attempt is allowed.
PNA Series Microwave Network Analyzers General Purpose Maintenance Procedures E8362B, E8363B, E8364B Firmware Upgrades Firmware Upgrades How to Check the Current Firmware Version 1. With the application running, click Network Analyzer Help About Network Analyzer A dialog box showing the current installed Application Code Version is displayed.
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers Operating System Recovery E8362B, E8363B, E8364B Operating System Recovery Recovering from Hard Disk Drive Problems If you suspect that you have a hard disk drive problem, go to the “Hard Drive Recovery”...
PNA Series Microwave Network Analyzers General Purpose Maintenance Procedures E8362B, E8363B, E8364B Correction Constants Correction Constants The analyzer stores many correction constants in non-volatile EEPROM memory. These constants enable the analyzer to produce accurate, leveled source signals and receive clean test signals.
Page 372
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers Correction Constants E8362B, E8363B, E8364B 8-22 Service Guide E8364-90026...
Page 373
Index Numerics part number 6-12 6-16 illustrated 7-33 removal and replacement 7-18 procedure 7-32 1284-C port illustrated 7-19 troubleshooting 4-45 rear panel connector 5-33 test nodes 4-37 A18 MA 20 troubleshooting 4-23 illustrated 6-21 troubleshooting 4-43 4-54 30-dB fixed attenuators operation A11 phase lock board illustrated...
Page 374
Index removal and replacement A30 channel B first converter illustrated 7-57 illustrated 6-21 illustrated 7-47 procedure 7-56 operation 5-21 procedure 7-46 troubleshooting 4-49 4-50 part number 6-20 A24 channel R2 detector A37 port 2 step attenuator illustrated 6-21 removal and replacement illustrated 6-25 6-27...
Page 375
Index location of 6-76 6-77 procedure 7-52 getting ready operation 5-33 troubleshooting 4-53 administrator password Agilent 7-66 part numbers 6-78 A6 SPAM board illustrated 6-13 6-17 Agilent removal and replacement administrator password 7-66 operation illustrated 7-63 support 2-12 analog 5-23 procedure 7-62 airline handling precautions...
Page 376
Index A45 and A46 reference interconnects 5-10 5-12 AUX I/O connector pin assignments 5-15 transfer switches 5-24 5-26 5-33 rear panel 5-12 A47 RF path coupler 5-26 rear panel replaceable parts A48 LO path coupler 5-26 illustrated 6-77 A49 IF multiplexer board part numbers 6-76 5-26...
Page 377
Index operation 5-18 part numbers 6-24 6-26 broadband failure 4-33 4-41 built-in tests removal and replacement bottom assembly, Option UNL/014 for commercial calibration, replaceable cables illustrated 7-59 Option 897 illustrated 6-33 6-35 procedure 7-58 for standards compliant part numbers 6-32 6-34 troubleshooting 4-51...
Page 378
Index illustrated 6-37 6-39 typical power output 4-48 protective end caps part numbers 6-36 6-38 channel A receiver module part numbers 6-80 troubleshooting 4-53 in top assembly, Option 080 repeatability check channel B illustrated 6-19 failure of 3-19 troubleshooting 4-31 4-48 part numbers 6-18...
Page 379
Index removal and replacement illustrated 6-45 6-47 illustrated 7-73 channel R1, Option 081 earth ground procedure 7-72 part number 6-44 electrostatic discharge, See ESD cover, removing channel R2 EMI/RFI covers, analyzer illustrated 6-21 supplies illustrated 6-75 operation 5-17 part numbers 6-82 part numbers 6-74...
Page 380
Index rear panel 5-10 channel R1 at 10 MHz 3-41 EXT 10 MHz REF OUT connector illustrated 6-21 at 3.8 MHz 3-42 rear panel 5-10 part number 6-20 frequency offset extender board functional description of removal and replacement measurement points illustrated 7-51 voltages...
Page 381
Index illustrated 6-11 part number 6-78 insertion loss check part numbers 6-10 removal and replacement failure of 3-15 front panel interface board illustrated 7-63 illustrated 6-11 test port cables 3-15 procedure 7-62 operation 5-31 installing replaceable parts a serial number 8-18 part number 6-10...
Page 382
Index LO offset test port cables 3-16 adjustment 3-45 maintenance jumpers, front-panel LO path coupler cleaning Option 014 removal and replacement preventive illustrated 6-29 6-31 illustrated 7-73 effect on error terms part number 6-28 6-30 procedure 7-72 manuals, printing copies LO path coupler, A48 max output power operation...
Page 383
Index processing subgroup 5-32 Option 080, frequency offset mode receiver group 5-21 N2721A software Option 081, transfer switch receiver mixers 5-21 tests 3-31 Option 099, firmware upgrade receiver modules 5-21 network analyzer, See analyzer Option 1CM, rack mount without receiver motherboard 5-23 noise floor, test 3-35...
Page 384
6-78 service 6-80 in internal analyzer 6-72 upgrade kits 6-81 in rear panel assembly 6-76 upgrade kits for E8362B 6-81 kits upgrade kits for E8363B 6-81 handle 6-82 upgrade kits for E8364B 6-81 30-dB rack mount...
Page 385
Index part numbers 6-78 part number 6-12 6-16 maximum test 3-32 in internal analyzer removal and replacement precautions handling airlines 3-24 illustrated 6-73 illustrated 7-21 pre-repair part numbers 6-72 procedure 7-20 procedures in rear panel assembly test nodes 4-37 caution about illustrated 6-77 phase stability check...
Page 386
Index GPIB 5-33 removal and replacement repair post-repair procedures 7-80 HANDLER I/O 5-12 procedure 7-52 pre-repair procedures 5-33 channel B procedures PARALLEL 5-33 illustrated 6-21 caution about PULSE IN 5-26 part number 6-20 warnings about RS-232 5-33 removal and replacement repairing SERIAL 5-33...
Page 387
Index illustrated 6-23 in internal analyzer A2 display assembly part numbers 6-22 illustrated 6-73 illustrated 7-13 replaceable cables, wrapped part numbers 6-72 procedure 7-12 in bottom assembly in rear panel assembly A20 LODA illustrated 6-67 illustrated 6-77 illustrated 7-39 part numbers 6-66 part numbers 6-76...
Page 388
Index procedure 7-56 assembly troubleshooting 4-23 A37 port 2 60-dB step cautions about attenuator list of procedures illustrated 7-57 sequence procedure 7-56 warnings about safety considerations A38 port 1 bias tee attenuators, 30 dB earth ground illustrated 7-59 illustrated 7-45 symbols procedure 7-58...
Page 389
Index options 8-16 splitter strap handle illustrated 6-21 analyzer SOMA 50 illustrated 6-21 operation illustrated 6-75 operation 5-11 part number 6-20 part number 6-74 part number 6-20 removal and replacement subgroup data acquisition and processing removal and replacement illustrated 7-37 5-29 illustrated 7-41...
Page 390
Index failure of 3-27 part number 6-68 6-70 frequency offset group (option getting ready cross talk test 3-37 080) 4-54 interpreting results 3-28 dynamic accuracy test 3-38 N2721A software 3-31 performing 3-23 maximum output power test performance 3-31 procedure 3-25 3-32 serial bus 4-35...
Page 391
4-53 upgrade kits 4-31 A32 channel R1 receiver part numbers 6-81 signal separation group 4-48 module 4-53 upgrade kits, E8362B source group 4-35 4-41 part numbers 6-81 A33 channel R2 receiver traces A, B, R1, R2 4-31 upgrade kits, E8363B...
Page 392
Index Option 016 troubleshooting 4-23 upgrade, reference transfer switch Option 081 upgrades warm up analyzer analyzer warning accessories warnings CD-RW drive for servicing warranty part numbers 6-82 service front panel one-year return to Agilent troubleshooting 4-17 2-12 port standard 2-12 front panel connector 5-31 Windows...