HP-IB Programming
OLA Specific Commands
:SENSe:POWer:REFerence:DISPlay
:SENSe[1|2]:POWer:REFerence:DISPlay
This command stores the value currently measured and displayed as
the new reference value (usually the source power). It is only
applicable to the applications
• Insertion Loss,
• PDL/Insertion Loss,
• Coupler Test,
• Directivity, and
• Powermeter.
Note: After starting the application wait ca. 2 s before issuing this
command.
1|2
1 Reads head A
2 Reads head B
Default is 1.
The message "Wrong application for this command" is returned, if
the present application is not one of those listed above.
The message "No head connected" is returned, if there is no optical
head attached to the specified channel.
:SENSe:POWer:REFerence:DISPlay?,
:SENSe:POWer:REFerence:DISPlay:HEAD?
OUTPUT 724;":SENS1:POW:REF:DISP"