Model HP E5574A Optical Loss Analyzer
Test
Test Description
No.
I.
Central Wavelength
performed at 1310 nm
Wavelength
performed at 1550 nm
Wavelength
II.
Output Power
fiber type 9/125 µm SM
performed at 1310 nm
performed at 1550 nm
III.
CW-Stability short term
fiber type 9/125 µm SM
performed at 1310 nm
15 minutes,
T = constant DIFF
Level variation peak-to-peak
performed at 1550 nm
15 minutes,
T = constant DIFF
Level variation peak-to-peak
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Appendix D. Performance Test
Performance Test Form Sheets
Performance Test for the HP 5574A
Minimum
Spec.
1290.00
1530.00
dBm
-10.00
-8.00 typ
-10.00
-8.00 typ
215
No. ____________ Date ___ / ___ / ___
Result
Maximum
Spec.
nm
nm
_________
1330.00
_________
1570.00
dBm
_________
_________
dBpp
_________
_________
Page 2 of 6
Measurement
Uncertainty
nm
__________ nm
dBm
_________ dBm
n/a
n/a
dBpp
__________ dB
0.02
0.02