HP-IB Programming
OLA Specific Commands
OUTPUT 724;":SCAN:RAT?"
ENTER 724;A$
:SENSe:CORRection:COLLect:ZERo
:SENSe[1|2]:CORRection:COLLect:ZERo
This command zeroes both optical heads (if attached). It is
applicable to all applications.
Note: The laser has to be switched off. The polarization controller
stops automatically. No other command will be executed until
zeroing is finished.
1|2 The channel parameter 1 or 2 may be present, but is ignored.
The message "Wrong application for this command" is returned, if
no application is active.
The message "No head connected" is returned, if there is no optical
head attached to the instrument.
:SENSe:CORRection:COLLect:ZERo?
OUTPUT 724;":SENS:CORR:COLL:ZER"
:SENSe:CORRection:COLLect:ZERo?
:SENSe[1|2]:CORRection:COLLect:ZERo?
This query returns the status of the most recent zero command.
1|2 The channel parameter 1 or 2 may be present, but is ignored.
A boolean value:
0
Success
1
Zeroing was not possible, one or both of the heads were
sensing
light.
:SENSe:CORRection:COLLect:ZERo
OUTPUT 724;":SENS:CORR:COLL:ZER?"
ENTER 724; A$