Figure D-4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Appendix D. Performance Test
Test IV. Linearity and Accuracy
5 Execute measurement:
a Press E
DIT
b Press E
XEC
c
The display shows RUNNING, and the remaining time,
counting down in seconds. The HP 8153A stops
automatically when the logging time is over, then displays
STABILITY.
6 Get measurement results, MIN/MAX readings:
a Press M
ORE
b Press E
DIT
c
Press N
EXT
d Note the displayed value in your test record.
Test IV. Linearity and Accuracy
The following steps describe the procedure for performing the
accuracy test for the HP E5574A / 81521B#001:
HP 8153A / 1
Standard Lightwave Multimeter
Standard Optical Head
Interface Module
HP 81533B / 1
Standard
Optical Interface
Optical Head
HP 81000AI

Accuracy Test Setup

202
to get back to STABILITY.
to execute datalogging.
to get SHOW.
to get MAXIMUM.
twice to get DIFF.
HP 8153A / 2
HP 8156 Attenuator
Standard Lightwave Multimeter
Depending on
Option of the
Attenuator
HP 81521B #001
Optical Head
Optical Adapter HP 81000AA
Standard Optical Head
Interface Module
HP 81533B / 2