Using the OLA as a Laser Source and Polarization Controller
Using the OLA as a Polarization Controller
How to Change the Scan Rate
If the automatic sweep of the polarization controller is on, the "Pol.
Controller" screen displays the current scan rate.

Polarization Controller Rate Settings Display

You can change the scan rate within a range of 2 to 6 sweeps per
minute.
To change the scan rate press
keypad to set the new value and confirm with
How to Terminate the Automatic Sweep
To stop the sweep do one of the following:
• Suspend the current application, e. g. by pressing A
• If you are using the "Return Loss" application, press
and
P
O
OLC
FF
return to the measurement screen.
• If you are using the "Powermeter" application, press
access the "Pol. Controller" screen and press
86
S
. R
. Use the rotary knob or the
C
ATE
. The automatic sweep will stop as soon as you
E
or E
.
NTER
NTER
.
PPL
S
ETTINGS
P
to
OLC
P
O
.
OLC
FF