5.4.16 Readoptionfunc - Edwards STP-iS1607 Series Instruction Manual

Turbomolecular pumps
Table of Contents

Advertisement

5.4.16 ReadOptionFunc

Function: Reads the setting value of the input operation port, warning function, and serial
communication time out.
Transmission frame:
PC→SIM Stx
0
0
SIM→PC
PC→SIM
SIM→PC Stx
0
0
*1
The HEX code of ASCII character ' = ' is "3D".
Parameter
1
Input operation port setting
*1
2
Enable/Disable setting
3
[System reservation]
4
Second Damage Limit Option
Enable/Disable setting
5
First Damage Limit Warning
Enable/Disable setting
Pump Runtime Over Warning
6
Enable/Disable setting
7
Pump Runtime Over Warning
Hours setting (×100 hours)
8
Imbalance Warning
Enable/Disable setting
Pump Overload Warning
9
Enable/Disable setting
Pump Overload Warning
10
Motor
11
Pump Overload Warning
Rotational speed setting
*1
Valid only with TMS specification
STP-iS1607 Series Turbomolecular Pump
*1
1
?
=
Etx LRC
*1
1
Sp
=
Parameter 1 to 13
Item
TMS Option
setting
current
Ack
Ack
Etx LRC
Data format
8-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
48-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
32-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
16-bits hexadecimal
coded ASCII
16-bits hexadecimal
coded ASCII
5
Remarks
0.1 % step
0.1 % step
MT-94E-001-C
Page 115

Advertisement

Table of Contents
loading

Table of Contents