Remote Operation
Power On Output State
You may also change the output state whether the output is enabled or disabled at
power on.
To change the power on at output state:
OUTP:PON:STAT [ON|OFF|1|0]
To check the setting:
OUTP:PON:STAT?
Reset
Resetting the unit puts certain features to a known state. These states are listed in
the table below.
To reset the unit:
*RST or SYST:RES
Table 4-8 Features Affected by Reset (*RST) Command
Feature
Voltage setpoint
Current setpoint
Power setpoint
Low voltage setpoint limit
High voltage setpoint limit
Low current setpoint limit
High current setpoint limit
Low power setpoint
High power setpoint
Over voltage protection
Under voltage protection
Over current protection
Under current protection
Over power protection
Under power protection
Fold shutdown protection
AC off shutdown recovery
Over temperature shutdown
recovery
Triggered voltage setpoint
Triggered current setpoint
4–24
Reset State (*RST)
0.0V
0.0A
103% of power rating
0.0V
103% of voltage rating
0.0A
103% of current rating
0% of power rating
103% of power rating
Disabled (0.0V)
Disabled (0.0V) and not shutdown when tripped
Disabled (0.0A) and not shutdown when tripped
Disabled (0.0A) and not shutdown when tripped
Disabled (0.0W) and not shutdown when tripped
Disabled (0.0W) and not shutdown when tripped
None and delay 0.5s
Autorecover
Latched
Disabled (Default)
Disabled (Default)
TM-XDOP-01XN