Configure Analog Control; Front Panel - Xantrex XDC 10-1200 Operating Manual

Xdc 6000 watt and 12000 watt series digital programmable dc power supply
Hide thumbs Also See for XDC 10-1200:
Table of Contents

Advertisement

Remote Operation

Configure Analog Control

You may configure the analog programming lines to work in either a 0 to 5V
range or a 0 to10V range. You may also select whether the voltage, current or both
are controlled with the programming lines. If you select only voltage or only
current, the other will be controlled via the front panel.

Front Panel

First, configure the analog programming lines.
Step # Do This
1
2
3
4
5
6
7
a. You can also press MENU 5 times.
Next select analog programming as the remote control interface.
Your options are
4–8
MENU
a
5 times
ENTER
ENTER
or
Select the input range.
ENTER
Analog V & I - voltage and current programmed via the analog interface
Analog V - voltage programmed via the analog interface; current set via front
panel
Analog I - current programmed via the analog interface; voltage set via front
panel
You Will See
ERROR MSGS
REMOTE CONFIG
RS-232 Cfg
Analog Cfg
Input 0-5V
0-5V or
0-10V
Setting is saved and menu is exited.
TM-XDOP-01XN