To set the Over-Voltage Protection:
The other protection options follow a similar procedure, but have 2 extra steps:
1. When you press PROT SET / ALARMS, the OVP prompt appears. Press
2. Protection options, other than OVP, prompt you with S/D if trip? N
These 2 extra steps are shown in the following example.
To set the Under-Voltage Protection:
TM-XDOP-01XN
Step # Do This
1
PROT SET
ALARMS
2
ENTER
3
Use the Voltage knob, the numeric, keypad,
or the arrow keys to enter a value. The value
must be between 0–103% of the unit's rated
voltage.
4
ENTER
PROT SET/ ALARMS repeatedly to cycle through the other protection
options.
Use the arrow keys to select Yes or No.
Yes shuts down the power supply if the protection limit is reached.
No issues a warning (a message in the Alarms menu) without shutting down
the supply.
Step # Do This
1
PROT SET
ALARMS
2
Press repeatedly until the
PROT SET
ALARMS
desired setting appears.
3
ENTER
4
Use the Voltage knob, the numeric
keypad, or the arrow keys to enter a
value (0–103% of the unit's rated
voltage).
5
ENTER
6
Use the arrow keys to select Yes or No.
For this example, Yes is selected.
7
ENTER
Power Supply Operation
You Will See
OVP SHUTDOWN
OVP SET 0V
OVP SET ####V
This setting is saved and the
display returns to its default
operating mode.
You Will See
OVP SET 0V
UVP SET 0V
UVP SET 0V
UVP SET ####V
S/D if trip? N
S/D if trip? Y
This setting is saved and the display
returns to its default operating mode.
3–11