Analog Programming Interface Current Readback Calibration; Analog Programming Interface 0-10V Range - Xantrex XDC 10-1200 Operating Manual

Xdc 6000 watt and 12000 watt series digital programmable dc power supply
Hide thumbs Also See for XDC 10-1200:
Table of Contents

Advertisement

Calibration (6000 Watt only)
7. Power supply calculates and stores calibration constants.
8. The menu will go to the 5V analog current readback calibration menu.

Analog Programming Interface Current Readback Calibration

1. Attach a DVM across the current readback lines, Pins B6 and B1(GND).
2. Enter 5V analog current readback calibration menu
3. Minimum calibration level
4. Enter voltage data
5. Maximum calibration level
6. Enter voltage data
7. Power supply calculates and stores calibration constants.
8. The menu will go to the 10V analog voltage programming calibration menu.

Analog Programming Interface 0-10V Range

The 0-10V range of the analog programming interface must be calibrated
separately. Follow the procedure exactly as for calibrating the 0-5V range, using
the corresponding 10V menu options. All the analog signals will be scaled by a
factor of 2.
B–10
Select ANLG I RB 5V from the calibration menu. Press ENTER.
You will be prompted to set up for analog current readback calibration. Press
ENTER when ready.
The power supply will automatically set the current readback lines to
approximately 10% of full scale.
Enter the voltage read from the external DVM. Press ENTER
The power supply will automatically set the current readback lines to
approximately 90% of full scale.
Enter the voltage at the current readback lines, read from the external DVM.
Press ENTER.
TM-XDOP-01XN