YOKOGAWA Admag TI Series User Manual page 135

Foundation fieldbus communication type
Hide thumbs Also See for Admag TI Series:
Table of Contents

Advertisement

Block
Relative
Name
Index
DTB
46
Diagnostic output
Menu path
Display
Device setup ► Diag/Service ► Verification ►(see below)
F
Diagnostic ► DTB ► Device Diagnostics ► Diagnostics/Alerts ► Maintenance ►
OUNDATION
Fieldbus
Verification ► (see below)
Block
Relative
Name
Index
DTB
48
Mode
DTB
49
Execute
DTB
50
VF No
DTB
47
VF target select
*1:
From the table below, select the output to execute the verification function.
Selection
F
OUNDATION
Display
Fieldbus
Zero
Zero
Measured value Measured value
Hold
Hold
*2:
From the table below, select the fluid state to execute the verification function.
Selection
Display
No flow
No flow
Flow
Flow
*3:
From the table below, select whether or not to execute the verification function. In the communication access of
F
Fieldbus, the communication access is set with the procedures of the interactive operation guide called
OUNDATION
DD Method.
Selection
Display
Not execute
Not execute
Execute
Execute
*4:
From the table below, select the execution result of the verification function.
Selection
Display
Factory
Factory
Previous
Previous
Present
Present
Parameter
Display
F
OUNDATION
Diagnostic Output
Parameter
Display
F
OUNDATION
Verification Mode
Verification Exe
Verification No
Verification Target
Process value
0% output
Output of
calculated value
(Undefined)
Fixes the last
valid value
before the
diagnosis starts.
F
OUNDATION
Fieldbus
Executes the verification function with the fluid not flowing.
Executes the verification function with the fluid flowing.
F
OUNDATION
Fieldbus
Does not execute the verification function.
Executes the verification function.
F
OUNDATION
Fieldbus
Displays the result obtained at shipment from the manufacturing
factory.
Displays the previous result.
Displays the result at this time.
<5. Functions>
Fieldbus
Specifies the output to execute
the verification function.
Fieldbus
Specifies the fluid status to execute
the verification function.
Specifies the execution of the
verification function.
Specifies the diagnosis result
display time.
Specifies the target for diagnosis.
Totalizer
Pulse output
Input 0
0 pps
(Output fixed)
Totalization of
Output of
calculated value
calculated value
(Undefined)
(Undefined)
Totalizes the
Fixes the last
last valid value
valid value
before the
before the
diagnosis starts.
diagnosis starts.
Description
Description
Description
134
Description
*1
Description
*2
*3
*4
*5
Frequency
output
0% output
Output of
calculated value
(Undefined)
Fixes the last
valid value
before the
diagnosis starts.
IM 01E21A02-03EN