ESD
Use a portable grounding mat if you cannot repair components at an ESD-protected
workstation. (Kodak's Customer Equipment Services Division can assist you in setting up
ESD-protected workstations.)
Use protective packaging when you transport components from one area to another.
Transparent anti static bags, available from a variety of manufactures, shield your
just-repaired components from further damage.

SAFETY WARNINGS

BE CAREFUL WHEN WORKING IN THE CASSETTE OPENER AREA. THE OPENER MOTOR AND
THE OPENER MECHANISM ARE VERY STRONG. THEY CANNOT BE MOVED MANUALLY. THEY
MAY SQUEEZE YOUR HAND AND TRAP YOU IF YOU TRY TO STOP THEM MANUALLY. NEVER
START THE CASSETTE OPENER MOTOR WHEN SOMEONE'S HANDS ARE IN THE CASSETTE
DISCONNECT THE POWER WHEN WORKING AT THE POWER SUPPLY. IT IS POSSIBLE TO
IF YOU TAKE OFF PANELS OR COVERS OR WIRE TIES TO GET EASIER ACCESS TO PARTS,
INSTALL THEM BEFORE YOU MOUNT THE SIDE PANELS WHEN THE SERVICE CALL OR THE
PM IS FINISHED. ENSURE THAT ALL PROTECTIVE GROUND WIRES ARE CONNECTED AGAIN.
BE CAREFUL WHEN WORKING ON CIRCUIT BOARDS A4 AND A8. THERE is 120 VAC ON
THESE BOARDS. BE ESPECIALLY CAREFUL WHEN MEASURING ON THE REAR SIDE OF THE
ENSURE THAT AFTER A SERVICE CALL OR PREVENTIVE MAINTENANCE THE INTERLOCK
6/1997
TOUCH THE MAINS VOLTAGE.
CIRCUIT BOARDS. AN ELECTRIC SHOCK MAY RESULT.
SWITCHES ARE FUNCTIONING.
1.
AREA.
2.
3.
4.
5.
II