X10 Inch V (Ml2000 P Only); Cassette Offset - Kodak Miniloader 2000 Service Manual

Table of Contents

Advertisement

PARAMETER

3. 8x10 inch V (ML2000 P only)

PURPOSE:
To enable the MULTILOADER to handle 8x10 inch Video Film Holder.
8x10" Video FILM HOLDER ........................................................................00h
15x30cm X-Ray CASSETTE (not available) ...........................................01h
1
.
Change 8x10"V / 15x30 cm.
Start the SERVICE PROGRAM
Select SERVICE MODE from the GLOBAL MENU .......................... press ENTER
ENTER SERVICE MODE MESSAGE is displayed ............................ press ENTER
UNIT DATA are displayed....................................................................... press ENTER
Select CHANGE ML2000 DATA from the MAIN MENU ............... press ENTER
Select CHANGE PARAMETER .............................................................. press ENTER
Select 8x10"V/15x30 cm .......................................................................... press ENTER
Key in the new value.................................................................................. press ENTER
Select STORE PARAMETER .................................................................. press ENTER
2
.
Exit the SERVICE PROGRAM.
Press BACKSPACE twice
Select QUIT ML2000 SERVICE MODE ............................................... press ENTER
Select Quit the program ............................................................................ press ENTER

4. CASSETTE OFFSET

PURPOSE:
To measure the correct CASSETTE LENGTH.
When the FUNCTION "CASSETTE LENGTH" is started, a CASSETTE with a known length is
put into the ML 2000. The program compares the known length against the amount of
CENTRING STEPS and calculates the necessary offset. From now on this offset is used to
determine the correct CASSETTE SIZE.
1
.
Start NEARLY EMPTY.
Start the SERVICE PROGRAM
Select SERVICE MODE from the GLOBAL MENU .......................... press ENTER
ENTER SERVICE MODE MESSAGE is displayed ............................ press ENTER
UNIT DATA are displayed....................................................................... press ENTER
Select CHANGE ML2000 DATA from the MAIN MENU ............... press ENTER
Select CHANGE PARAMETER .............................................................. press ENTER
Select CASSETTE LENGTH ................................................................. press ENTER
Key in the measured CASSETTE LENGTH..................................... press ENTER
Select STORE PARAMETER .................................................................. press ENTER
6/1997
Use the smallest CASSETTE available. (18x24cm)
NOTE
4-4
SM 3477
KODAK AG, Stuttgart