INSTALLATION INSTRUCTIONS
Kodak MINILOADER 2 PLUS LOW VOLTAGE
Use this Publication for:-
MINILOADER 2 Plus Low Voltage Model Stand-Alone (SV code 3419), MINILOADER 2 Plus Low Voltage Model
M35-M (SV code 3420) and MINILOADER 2 Plus Low Voltage Model 480RA (SV code 3421).
HEALTH SCIENCES DIVISION
© Kodak Ltd. 1995
for the
Publication No. II 3420
January 1995

Advertisement

loading

Summary of Contents for Kodak HEALTH SCIENCES MINILOADER 2 Plus Low Voltage Stand-Alone

  • Page 1 January 1995 INSTALLATION INSTRUCTIONS for the Kodak MINILOADER 2 PLUS LOW VOLTAGE Use this Publication for:- MINILOADER 2 Plus Low Voltage Model Stand-Alone (SV code 3419), MINILOADER 2 Plus Low Voltage Model M35-M (SV code 3420) and MINILOADER 2 Plus Low Voltage Model 480RA (SV code 3421).
  • Page 2 Kodak reserves the right to change this information without notice, and makes no warranty, express or implied, with respect to this information. Kodak shall not be liable for any loss or damage, including consequential or special damages, resulting from the use of this information, even if loss or damage is caused by Kodak's negligence or other fault.
  • Page 3: Table Of Contents

    FOR M35-M VERSION UNPACKING AND INSTALLATION INSTRUCTIONS FOR 480RA VERSION UNPACKING AND INSTALLATION INSTRUCTIONS FOR STAND-ALONE VERSION MAMMOGRAPHY CASSETTE PREPARATION VIDEO CASSETTE PREPARATION INSTALLING THE HUMIDIFIER INSTALLATION CHECK LIST APPENDIX "A" - SUITABLE STANDS January 1995 Page II-3 © Kodak Ltd.
  • Page 4: Unpacking And Installation Instructions For M35-M Version

    Remove the REAR PANEL, disconnecting the EXHAUST HOSE from the EXHAUST ADAPTOR. Remove the EXHAUST ADAPTOR, and the PANEL FIXING SCREWS and RETAINING WASHERS from the REAR PANEL and keep. Discard all PANELS. © Kodak Ltd. Page II-4 January 1995...
  • Page 5 PROCESSOR under the LEVELLING FEET, and level the PLINTH leaving a gap of approximately 25 mm between the floor and the bottom of the PLINTH. See FIGURE 2. Figure 2 Processor plinth 25 mm January 1995 Page II-5 © Kodak Ltd.
  • Page 6 Remove the LID and lift the MINILOADER onto its STAND. The front of the MINILOADER is at the front of the STAND. IMPORTANT - ONLY LIFT THE MINILOADER BY THE FRAME AS SHOWN IN FIGURE 4. © Kodak Ltd. Page II-6 January 1995...
  • Page 7 The packing BLOCKS under the TWO SUCKER BARS. c) The protection PACKING from the film exit CHUTE. d) The TIE-WRAP from the MAGAZINE INJECTOR BAR. e) The TIE from the CASSETTE INJECTOR. January 1995 Page II-7 © Kodak Ltd.
  • Page 8 SCREWS etc. Connect the EARTH LEAD from the PANEL to the M35-M ELECTRICAL BOX as shown in FIGURE 6. Refit the STAND-BY/RUN SWITCH. Reconnect the M35-M EXHAUST HOSE to the EXHAUST ADAPTOR. Figure 6 New rear panel Electrical box Earth lead © Kodak Ltd. Page II-8 January 1995...
  • Page 9 Complete all the PLUMBING and WIRING to the PROCESSOR. See FIGURE 8 for the purpose of the HOLES in the REAR PANEL. Figure 8 Stand-by/Run switch Exhaust Replenisher inlets Processor power cable Drain outlets Water inlet January 1995 Page II-9 © Kodak Ltd.
  • Page 10 Fit the FAN GRID, using the original SCREWS, but adding the M4 STAR WASHERS. IMPORTANT - THE PAINT MUST BE REMOVED FROM THE FRAME WHERE THE STAR WASHERS CONTACT IT. See FIGURE 10. New Star Washers Screw Grid Screw Screw Grid Remove paint here Figure 10 © Kodak Ltd. Page II-10 January 1995...
  • Page 11 Existing screw New metal bracket (30026254) Behind original clip Figure 12 Countersunk Screw 3 Countersunk Screw 2 New clear Nuts (30015612) panel (30026205) Screw (30015691) Washer Behind original clip Behind original clip (30015569) January 1995 Page II-11 © Kodak Ltd.
  • Page 12 MANUAL. IMPORTANT, THE M35-M MUST BE INSTALLED IN ACCORDANCE WITH THE M35-M INSTALLATION MANUAL AND ADDITIONALLY IN ACCORDANCE WITH LOCAL PRACTICES. Then go to, “INSTALLING THE HUMIDIFIER” on page II-22 of this PUBLICATION. © Kodak Ltd. Page II-12 January 1995...
  • Page 13: Unpacking And Installation Instructions For 480Ra Version

    If the STAND is metal, an earth connection must be made between the STAND and the MINILOADER.. IMPORTANT - ENSURE THE PAINT IS REMOVED WHERE THE EARTH CONNECTIONS ARE MADE. Remove the left hand SIDE PANEL from the MINILOADER. January 1995 Page II-13 © Kodak Ltd.
  • Page 14 30, and the two THIN M4 NUTS as shown below in figure 15. Place one of the M4 LARGE DIAMETER WASHERS on each SCREW. Figure 15 Miniloader Processor Processor entry slot Screw Washers Thin nut © Kodak Ltd. Page II-14 January 1995...
  • Page 15 VERTICAL 1.5 mm ALIGNMENT When both horizontal and vertical alignments are correct, tighten the M4 NUTS. Go to, “INSTALLING THE HUMIDIFIER” on page II-22 of this publication to complete the installation.. January 1995 Page II-15 © Kodak Ltd.
  • Page 16: Unpacking And Installation Instructions For Stand-Alone Version

    !! CAUTION TWO PEOPLE ARE REQUIRED TO LIFT THE MINILOADER !! 690 mm centres Figure 18 Drill 4 HOLES 10 mm diameter in the STAND or BENCH. Feed Receiving 660 mm 580 mm end. centres 770 mm © Kodak Ltd. Page II-16 January 1995...
  • Page 17 24 x 30 RECEIVING MAGAZINE onto the outside of the MAGAZINE over the existing LABEL. Do not fix the, "24 x 30" LABEL or the REFLECTIVE PATCH to the MAGAZINE. January 1995 Page II-17 © Kodak Ltd.
  • Page 18 Unscrew the PHOTOCELL and move it to the opposite end of the mounting bracket. See figure 21. Remember to adjust the PHOTOCELL to give correct MAGAZINE FULL detection (65 FILMS maximum) as you complete the installation. Original position of FC7 New position of photocell Figure 21 © Kodak Ltd. Page II-18 January 1995...
  • Page 19 ALL FILM SIZES COLLECTED 18 x 24 & 24 x 30 FILMS IN THE SAME RECEIVING COLLECTED MAGAZINE SEPARATELY Go to, “INSTALLING THE HUMIDIFIER” on page II-22 of this publication to complete the installation. January 1995 Page II-19 © Kodak Ltd.
  • Page 20: Mammography Cassette Preparation

    If DIPSWITCH 6 on PCB 401 is turned on, no MAMMOGRAPHY CASSETTES should have a second (TYPE 2) patch. PATCH 1 This is the TYPE 2 position. Figure 23 Figure 24 PATCH 2 PATCH 3 © Kodak Ltd. Page II-20 January 1995...
  • Page 21: Video Cassette Preparation

    26. There are ten FELT PADS in a kit, and they are supplied with installation instructions. The part number of the FELT PAD KIT is 9166451. Figure 26 January 1995 Page II-21 © Kodak Ltd.
  • Page 22: Installing The Humidifier

    The position of the services to the HUMIDIFIER depend on the MODEL concerned, see the FIGURES below and opposite. If it is more convenient, the alternative DRAIN POSITION can be used. See DRAWING opposite. MODEL M35-M MODEL 480RA INLET INLET DRAIN DRAIN © Kodak Ltd. Page II-22 January 1995...
  • Page 23 For full details of the HUMIDIFIER, see the, “HUMIDIFIER SERVICE MANUAL” Publication number XP-3239-2. This Publication includes the SERVICE DETAILS, THEORY GUIDE, CIRCUIT DIAGRAMS and PARTS LIST. Go to the, “INSTALLATION CHECK LIST” on page II-24 of this PUBLICATION. January 1995 Page II-23 © Kodak Ltd.
  • Page 24 See the TABLE and DRAWING below. MEASURED VOLTAGE TAPPING CONNECTION 95 - 104 TB2-2 105 - 114 TB2-3 115 - 126 TB2-4 210 - 225 TB2-5 226- 235 TB2-6 236 - 252 TB2-7 © Kodak Ltd. Page II-24 January 1995...
  • Page 25 FILM separation in the SUPPLY MAGAZINE, correct insertion of FILMS into the CASSETTES and correct delivery of FILM into the CHUTE (or RECEIVING MAGAZINE, depending on version of MINILOADER). Make any adjustments necessary. January 1995 Page II-25 © Kodak Ltd.
  • Page 26 Air blow cassette opening 24 x 30 026 - 032 Display brightness 1 - 4 Multiple film check 18 x 24 149 - 160 Multiple film check 24 x 30 143 - 153 © Kodak Ltd. Page II-26 January 1995...
  • Page 27 Fill in the INSTALLATION REPLY FORM, and note the meter (counter) reading. Make sure the correct SERVICE CODE is used when feeding back the installation to SCAN. SERVICE CODE MODEL 3419 STAND-ALONE 3420 M35-M 3421 480RA January 1995 Page II-27 © Kodak Ltd.
  • Page 28 16) in the TOP to bolt the MINILOADER to the STAND. FRONT VIEW OVERALL HEIGHT THESE CASTORS MUST BE RECESSED 4 TWIN WHEEL HEIGHT TO ALLOW THE ADJUSTABLE CASTORS STAND TO WHEEL UP WITH BRAKED PLATE TO THE PROCESSOR FIXING © Kodak Ltd. Page II-28 January 1995...
  • Page 29 50/60 CUTOUT FOR SERVICES KICKING WELL 50/60 DEPENDING ON CASTOR WIDTH CONSTRUCTION IN 25 mm MELAMINE FACED LAMINATED BOARD ALL DIMENSIONS IN MILLIMETRES - NOT TO SCALE - DO NOT SCALE FROM DRAWING January 1995 Page II-29 © Kodak Ltd.
  • Page 30 The CASTORS should be adjustable for height and should be capable of carrying a load of 50 Kg each. All construction should be painted steel. This design is not suitable for a PROCESSOR INTERFACE MINILOADER. © Kodak Ltd. Page II-30 January 1995...
  • Page 31 INSTALLATION INSTRUCTIONS II 3420 THIS PAGE IS INTENTIONALLY LEFT BLANK January 1995 Page II-31 © Kodak Ltd.
  • Page 32 HEALTH SCIENCES DIVISION © Kodak Ltd., Customer Equipment Services Division, Swallowdale Lane, Hemel Hempstead, Hertfordshire, England.