Mounting and sensor wiring
April 2023
4. Lower the transmitter, sliding it onto the notches on the feedthrough.
5. Replace the clamping ring on the feed through. Tighten the screw to 28 in lbf (3.16 N
2.6 

Rotating the display orientation

The user interface orientation for the transmitter can rotate 360° in 90° increments by
software selection.
Using the display, select Menu → Configuration → Display Settings → Rotation.
2.7 
Rotate the sensor wiring junction box on a
remote-mount transmitter (optional)
In remote-mount installations, you can rotate the sensor wiring junction box on the
transmitter up to plus or minus 180 degrees.
Procedure
1. Using a 4 mm hex key, loosen and remove the clamp securing the sensor wiring
18
 CAUTION
Do not rotate the housing more than 360°. Excessive rotation can damage the
wiring and cause measurement error or flow meter failure.
m)– 32 in lbf (3.62 N m).
 CAUTION
Ensure that the connection between the transmitter and the sensor is moisture-
proof. Inspect and grease all gaskets and O-rings. Moisture in the electronics can
cause measurement error or flow meter failure.
junction box in place.