8 Part replacement instructions
6. Lift off the i4000 board with plate and fan.
7. Put the new i4000 board with plate and fan in place.
8. Insert and tighten the screws.
9. Connect the connectors and cables. Follow the markings on the connectors and the
circuit board.
10. Attach the laser assembly.
11. Align the laser lights.
12. For calibrations, see

8.3.5 Replacing the carriage cable set

Prerequisites: The horizontal lasers assembly and the i4000 power supply board have
been removed.
NOTICE! The carriage right and left cover and upper shelf right and left cover
must be removed before this procedure.
NOTICE! Take pictures of the carriage cable set before you remove it. This helps
you in installing the new set.
1. Disconnect the Ethernet cable.