7. Remove the two screws that hold the rotating unit.
8. Slide the rotating unit away from the column.
9. Lift the rotating unit off and lay it on the floor.
10. Remove the finger guard of the main cable.
11. Turn up the metal plates (four pcs) to release the cable.
12. Put the new cable coil to its place.
ORTHOPANTOMOGRAPH
NOTICE! It is recommended that two men lift the rotating unit.
NOTICE! The correct position of the cable is indicated with markings on the
cable.
OP 3D
8 Part replacement instructions
127