7 Troubleshooting
Error code
E21202
E21203
E21204
E21205
E21206
E21207
E21208
E21209
E21210
E21211
E21212
E21213
E22002
E22202
82
Problem
Goto Fixed - no opto found
Calib ABS - opto hit invalid ADC
value
Calib ABS - opto hit ADC value
change too small
Calib ABS - opto hit too early
Calib ABS verify - too large error See E09017 Resistive sensor missing.
Route - no valid route found
Goto ADC - Invalid start position Software error. Take logs and boot the
Goto ADC - Opto hit too fast
Goto ADC - Opto edge missing
Goto ADC - Edges end not found 1. Check the location of the resistive
Goto ADC - Adc value not
changing
Goto ADC - Adc value changing
to wrong direction
Route - waypoints on invalid
state
Goto Fixed - no opto found
Solution
1. Check connections:
* J3005, J1026
If fail (open) => replace cable pivot
sensor movement cable
2. Replace opto fork
3. Replace opto ring if any of the
teeth are broken
4. Replace i3000.
See E09017 Resistive sensor missing.
See E09017 Resistive sensor missing.
See E09017 Resistive sensor missing.
Software error. Take logs and boot the
device.
device.
See E26203 Goto ADC -Opto hit too
fast.
See E26204.
See E26205.
position sensor.
If fail => replace sensor and
reposition it.
See E09017.
1. Check connections:
* J2002, J1012
If fail => replace rotating high torq
motor cable
* J2001, J1013
If fail => replace rotating sensor
cable.
Software error. Take logs and boot the
device.
1. Check connections:
* J3004, J1025