NOTICE! To replace the patient positioning panel, you need to replace the whole
upper lower shelf cover.
Replace the board by following the instructions in
and lower

8.3.11 Replacing the collimator

NOTICE! You must remove the tube head front and back covers before this
procedure.
Required tools: Allen key 3mm
1. Detach the three cables from their connectors from the top of the collimator.
ORTHOPANTOMOGRAPH
cover.
OP 3D
8 Part replacement instructions
Removing the lower shelf upper
137