Replacing The I3000 Main Control Board - KaVo ORTHOPANTOMOGRAPH OP 3D Service Manual

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

1. Detach the connectors.
2. Remove the screws (seven pcs).
3. Lift the board off.
4. Place the new i2000 board onto its place.
5. Attach the screws (seven pcs) and tighten them.
6. Attach the cables and connectors.
7. For device calibrations, see

8.3.8 Replacing the i3000 main control board

NOTICE! You must remove the upper shelf right and left cover before this
procedure.
Required tools: screwdriver
1. Remove the connectors from both ends of the i3000 main control board.
ORTHOPANTOMOGRAPH
When to calibrate the
OP 3D
8 Part replacement instructions
device.
129