4 Testing | 4.1 Test of the SURGmatic S201 L

4 Testing

4.1 Test of the SURGmatic S201 L

4.1.1 Test set-up

① Test device 2182 KL
(Mat. no. 0.411.8540)
② Leakage air hose
③ Motor hose

4.1.2 Test instructions

Note
The following evidence must be documented:
- Customer's address and repair dates
- Material number or sales number of the product
- Serial number of the product
- Spare parts material numbers
- Name, date, signature or personnel number of repair/test engineer
Requirement
- Amount of cooling air: 28 ± 4 Nl/min
- Spray water pressure: 2 ± 0.1 bar
- Drive speed: max. 40,000 min
- The pressures are to be set on the test device.
④ Pressure gauge
(Mat. no. 1.003.1050)
⑤ Motor
-1
16/19