KaVo ORTHOPANTOMOGRAPH OP 3D Service Manual page 71

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

Error code
E10203
E10204
E10205
ORTHOPANTOMOGRAPH
Problem
kV feedback out of range
Tube fail
Calibration feedback not
increasing
OP 3D
Solution
2. Check connections: J4019, J1005,
J7000 (pan-THA)
3. Re-run THA calibrations
4. Replace i4000
5. Re-run THA calibrations
6. Replace pan THA
7. Re-run THA calibrations
8. Replace main cable and twister
cable assembly
9. Re-run THA calibrations.
1. Re-run THA calibrations
2. Check connections: J4005, J4019,
J1004, J1005, J7000 and J 7002 (pan
THA)
3. Re-run THA calibrations
4. Replace i4000
5. Re-run THA calibrations
6. Replace pan-THA
7. Re-run THA calibrations
8. Replace main cable and twister
cable assembly
9. Re-run THA calibrations.
1. Re-run THA calibrations
2. Check connections: J4005, J4019,
J1004, J1005, J7000 and J7002 (pan
THA)
3. Re-run THA calibrations
4. Replace i4000
5. Re-run THA calibrations
6. Replace pan THA
7. Re-run THA calibrations
8. Replace main cable and twister
cable assembly
9. Re-run THA calibrations.
1. Check connections: J4005, J4019,
J1004, J1005, J7000 and J 7002 (pan
THA)
2. Re-run THA calibrations
3. Replace i4000
4. Re-run THA calibrations
5. Replace pan THA
7 Troubleshooting