7 Troubleshooting
Error code
E23203
E23204
E23205
E23206
E23207
E23208
E23209
E23210
E23211
E23212
E23213
E26202
84
Problem
Calib ABS - opto hit invalid ADC
value
Calib ABS - opto hit ADC value
change too small
Calib ABS - opto hit too early
Calib ABS verify - too large error See E09019
Route - no valid route found
Goto ADC - Invalid start position Software error. Take logs and boot the
Goto ADC - Opto hit too fast
Goto ADC - Opto edge missing
Goto ADC - Edges end not found 1. Check the location of the resistive
Goto ADC - Adc value not
changing
Goto ADC - Adc value changing
to wrong direction
Position fail
Solution
If fail (open) => Replace rotating
sensor cable
2. Replace opto fork
3. Replace belt wheel if any of teeth
are broken
4. Replace i2000.
See E09019
See E09019
See E09019
Software error. Take logs and boot the
device.
device.
See E28203
See E28204
See E28205
position sensor.
If fail => Replace sensor and
reposition it.
See E09019
1. Check connections:
* J2002, J1012
If fail => Replace rotating high torq
motor cable
* J2001, J1013
If fail => Replace rotating sensor
cable.
1. Pushed/blocked externally.
2. See additional errors:
* E09011 Opto missing
* E09017 Resistive sensor missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
ORTHOPANTOMOGRAPH
OP 3D